Prolactin Monomeric Polyethylene Glycol Measurement Method and Study of Reference Value Verification
이용수 0
- 영문명
- 발행기관
- 대한핵의학기술학회
- 저자명
- Dong Hyuk Ha Hwa-Jin Ryu Hyun-Su Cho Sun-Young Shin
- 간행물 정보
- 『핵의학기술』Vol.27 No.2, 133~136쪽, 전체 4쪽
- 주제분류
- 의약학 > 방사선과학
- 파일형태
- 발행일자
- 2023.11.30
4,000원
구매일시로부터 72시간 이내에 다운로드 가능합니다.
이 학술논문 정보는 (주)교보문고와 각 발행기관 사이에 저작물 이용 계약이 체결된 것으로, 교보문고를 통해 제공되고 있습니다.

국문 초록
Purpose: Prolactin in the blood is separated into three types, and over 90% of prolactin presents as a double monomer (23 KDa). Rarely, it can exist in the size of big prolactin (150 KDa), which is called macroprolactin and is known as an autoantibody complex. When macroprolactin accounts for more than 60% of prolactin in the blood, it is called macroprolactinemia. The presence of such macroprolactin was first reported in a patient with hyperprolactinemia but without typical symptoms. Macroprolactinemia is emerging as an important cause of idiopathic hyperprolactinemia. The polyethylene glycol (PEG) precipitation method using the property of precipitating large-molecular-weight proteins is simple and recently has been widely used as a screening test. The results are in good agreement with the results of gel chromatography. The purpose of this study was to confirm the measurement method and reference value verification of monomeric prolactin in blood prolactin using the PEG precipitation method.
Materials and Methods: For 40 examinees who visited the Gangnam Center of Seoul National University Hospital in 2021, the prolactin level was verified using radioimmunoassay (RIA). For macroprolactinemia PEG precipitation method, 25% PEG (molecular weight 6000kDa) solution and serum were mixed in equal amounts in a test tube, then left at room temperature for 20 minutes and centrifuged at 4℃ for 30 minutes (1500g). The prolactin level was measured in the supernatant.
Results : After confirming that more than 90% of the 40 tested samples within the reference range <25 ng/mL, the same value as the reference value for prolactin was applied. Since the concentration of monomeric prolactin in serum from which macroprolactin has been removed from blood is diluted 1:1 with PEG, our laboratory is currently reporting the result by multiplying the result by a dilution factor of 2.
Conclusion: Radioimmunoassay using PEG precipitation method using the property of precipitating large molecular weight proteins is simple and effective for quantitative measurement of monomeric prolactin in blood prolactin.
영문 초록
목차
서론 및 목적
대상 및 방법
결과
고찰 및 결론
REFERENCES
해당간행물 수록 논문
- Study on the PET image quality according to various scintillation detectors based on the Monte Carlo simulation
- Prolactin Monomeric Polyethylene Glycol Measurement Method and Study of Reference Value Verification
- Synthesis of Novel 18F-Labeled-Nitroimidazole-Based Imaging Agents for Hypoxia: Recent Advances
- Calculation of the Least Significant Change Value of Bone Densitometry Using a Dual-Energy X-ray Absorptiometry System
- Femoral Mass in Bone Mineral Densitometry
- L-spine Bone SPECT/CT에서 획득된 저선량 CT 영상을 이용한 용적 골밀도 결과의 유용성
- 68Ga 표지 PET/CT 검사의 최적화된 매개변수에 대한 연구
참고문헌
관련논문
의약학 > 방사선과학분야 BEST
- 민간의료보험 해약 의향 영향 요인: 한국 의료 패널 2019년 2기 자료를 이용하여
- 고준위방사성폐기물 관리와 관련된 사회안전망 구축에 관한 연구
- 임상에서 사용되는 온장고 온도설정 및 한계점에 대한 고찰
의약학 > 방사선과학분야 NEW
- 무증상 치밀 유방 여성에서 인공지능 기반 컴퓨터 보조 진단 장치의 유용성 연구
- 방사선치료 임상실습 환경에 따른 방사선(학)과 학생의 학습 효과 분석
- 딥러닝을 활용한 한국형 유방초음파 영상 정상-종괴 구분 연구
최근 이용한 논문
교보eBook 첫 방문을 환영 합니다!
신규가입 혜택 지급이 완료 되었습니다.
바로 사용 가능한 교보e캐시 1,000원 (유효기간 7일)
지금 바로 교보eBook의 다양한 콘텐츠를 이용해 보세요!