학술논문
L-spine Bone SPECT/CT에서 획득된 저선량 CT 영상을 이용한 용적 골밀도 결과의 유용성
이용수 0
- 영문명
- Usefulness of volumetric BMD measurement by using low dose CT image acquired on L-spine Bone SPECT/CT
- 발행기관
- 대한핵의학기술학회
- 저자명
- 고현수(Hyun Soo Ko) 박순기(Soon Ki Park) 김은혜(Eun Hye Kim) 최종숙(Jong Sook Choi) 정우영(Woo Young Jung) 이동윤(Dong Yun Lee)
- 간행물 정보
- 『핵의학기술』Vol.27 No.2, 99~109쪽, 전체 11쪽
- 주제분류
- 의약학 > 방사선과학
- 파일형태
- 발행일자
- 2023.11.30
4,120원
구매일시로부터 72시간 이내에 다운로드 가능합니다.
이 학술논문 정보는 (주)교보문고와 각 발행기관 사이에 저작물 이용 계약이 체결된 것으로, 교보문고를 통해 제공되고 있습니다.

국문 초록
Purpose: CT scan makes up for the weak point of the nuclear medicine image having a low resolution and also were used for attenuation correction on image reconstruction. Recently, many studies try to make use of CT images additionally, one of them is to measure the bone mineral density(BMD) using Quantitative CT(QCT) software. BMD exams are performed to scan lumbar and femur with DXA(Dual-Energy X-Ray Absorptiometry) in order to diagnose bone disease such as osteopenia, osteoporosis. The purpose of this study is to identify the usefulness of QCT_BMD analyzed with low dose CT images on L-spine Bone SPECT/CT comparing with DXA_BMD.
Materials and Methods: Fifty five women over 50 years old (mean 66.4 ± 9.1) who took the both examinations(L-spine Bone SPECT/CT with SIEMENS Intevo 16 and DXA scan with GE Lunar prodigy advance) within 90 days from April 2017 to July 2022, BMD, T-score and disease classification were analyzed. Three-dimensional BMD was analyzed with low dose CT images acquired on L-spine Bone SPECT/CT scan on Mindways QCT PROTM software and two-dimensional BMD was analyzed on DXA scan. Basically, Lumbar 1-4 were analyzed and the patients who has lesion or spine implants on L-spine were excluded for this study. Pearson's correlation analysis was performed in BMD and T-score, chi-square test was performed in disease classification between QCT and DXA.
Results: On 55 patients, the minimum of QCT_BMD was 18.10, maximum was 166.50, average was 82.71 ± 31.5 mg/cm3. And the minimum of DXA-BMD was 0.540, maximum was 1.302, average was 0.902 ± 0.201 g/cm2, respectively. The result shows a strong statistical correlation between QCT_BMD and DXA_BMD(p<0.001, r=0.76). The minimum of QCT_T-score was -5.7, maximum was -0.1, average was -3.2 ± 1.3 and the minimum of DXA_T-score was -5.0, maximum was 1.7, average was -2.0 ± 1.3, respectively. The result shows a statistical correlation between QCT T-score and DXA T-score (p<0.001, r=0.66). On the disease classification, normal was 5, osteopenia was 25, osteoporosis was 25 in QCT and normal was 10, osteopenia was 25, osteoporosis was 20 in DXA. There was under-estimation of bone decrease relatively on DXA than QCT, but there was no significant differences statistically by chi-square test between QCT and DXA.
Conclusion: Through this study, we could identify that the QCT measurement with low dose CT images QCT from L-Spine Bone SPECT/CT was reliable because of a strong statistical correlation between QCT_BMD and DXA_BMD. Bone SPECT/CT scan can provide three-dimensional information also BMD measurement with CT images. In the future, rather than various exams such as CT, BMD, Bone scan are performed, it will be possible to provide multipurpose information via only SPECT/CT scan. In addition, it will be very helpful clinically in the sense that we can provide a diagnosis of potential osteoporosis, especially in middleaged patients.
영문 초록
목차
서론
실험 대상 및 방법
결과
고찰
결론
REFERENCES
키워드
해당간행물 수록 논문
- Femoral Mass in Bone Mineral Densitometry
- Calculation of the Least Significant Change Value of Bone Densitometry Using a Dual-Energy X-ray Absorptiometry System
- Synthesis of Novel 18F-Labeled-Nitroimidazole-Based Imaging Agents for Hypoxia: Recent Advances
- L-spine Bone SPECT/CT에서 획득된 저선량 CT 영상을 이용한 용적 골밀도 결과의 유용성
- 68Ga 표지 PET/CT 검사의 최적화된 매개변수에 대한 연구
- Study on the PET image quality according to various scintillation detectors based on the Monte Carlo simulation
- Prolactin Monomeric Polyethylene Glycol Measurement Method and Study of Reference Value Verification
참고문헌
관련논문
의약학 > 방사선과학분야 BEST
- 민간의료보험 해약 의향 영향 요인: 한국 의료 패널 2019년 2기 자료를 이용하여
- 방사선학과 학생 실습환경에 관한 연구: 실습실 내 미세먼지, 휘발성유기화합물 위해도를 중심으로
- 임상에서 사용되는 온장고 온도설정 및 한계점에 대한 고찰
의약학 > 방사선과학분야 NEW
- Femoral Mass in Bone Mineral Densitometry
- Calculation of the Least Significant Change Value of Bone Densitometry Using a Dual-Energy X-ray Absorptiometry System
- Synthesis of Novel 18F-Labeled-Nitroimidazole-Based Imaging Agents for Hypoxia: Recent Advances
최근 이용한 논문
교보eBook 첫 방문을 환영 합니다!
신규가입 혜택 지급이 완료 되었습니다.
바로 사용 가능한 교보e캐시 1,000원 (유효기간 7일)
지금 바로 교보eBook의 다양한 콘텐츠를 이용해 보세요!