- 영문명
- The Usefulness of Prospectively ECG Gated High-pitch Cardiac Scan for CT Coronary Angiography Using Dual Source CT
- 발행기관
- 대한CT영상기술학회
- 저자명
- 이대규(Dae Kyu Lee) 김문찬(Moon Chan Kim) 방성호(Sung Ho Bang) 한동균(Dong Kyoon Han)
- 간행물 정보
- 『대한CT영상기술학회지』대한전산화단층기술학회지 제12권 제1호, 25~33쪽, 전체 9쪽
- 주제분류
- 의약학 > 방사선과학
- 파일형태
- 발행일자
- 2010.04.30
4,000원
구매일시로부터 72시간 이내에 다운로드 가능합니다.
이 학술논문 정보는 (주)교보문고와 각 발행기관 사이에 저작물 이용 계약이 체결된 것으로, 교보문고를 통해 제공되고 있습니다.
국문 초록
영문 초록
Purpose
To evaluate the usefulness of prospectively ECG gated high-pitch cardiac scan for CT coronary angiography(CTCA) using dual source CT by analyzing the dose effect and image quality for CTCA protocols.
Materials and methods
CTCA was performed by using 5 current clinical protocols to evaluate effective dose and organ dose in prima1y beam area with anthropomorphic female phantom and glass dosimetric system. Also, ACR phantom was used to analyze image quality at 128 channel multi-detector CT (scan parameters: 320 mAs at 120 kVp, collimation: 128 × 0.6 mm, scan range :137 mm, reconstructed slice thickness 0.6 mm, increment 0.4 mm). Heart beat rate of phantom was maintained 60 bpm in ECG gating. In a retrospectively ECG gated helical scan mode was done with 280 msecgantry rotation time, 0.23 pitch. Peak dose was used without cardiac dose modulation (pulsing off mode). Peak dose was adopted in range of 60%~80% of R-R interval and 25% reduction in others with cardiac dose modulation (pulsing manual mode). And peak dose was adopted in range of 60%~80% of R-R interval and 4% reduction in others with cardiac dose modulation (pulsing MinDose-manual mode). In a prospectively ECG gated sequential scan mode data were acquired at 70% R-R interval (middle diastolic phase in cardiac cycle), and 170 msec gantry rotation time was used. In a prospectively ECG gated high-pitch cardiac scan mode data were acquired from 60% R-R interval, and 3.4 pitch was used. Analysis of image quality was clone by evaluating objective noise, uniformity, high contrast resolution, and linearity.
Results
The mean effective dose for pulsing off mode without cardiac dose modulation in a retrospectively ECG gated helical scan was 20.8 mSv, and die mean organ dose of heart was 116.7 mGy. Pulsing manual mode showed 49% reduction, compared with that of pulsing off mode. And pulsing MinDose-manual mode showed 60% reduction compared with that of pulsing off mode. And at die adaptive sequential scan mode under prospective ECG gating die mean effective dose was 2.7 mSv, and die mean organ dose of heart was 17.5 mGy, d1is represents an 86% reduction compared with that of pulsing off mode. In the high pitch cardiac scan under prospective ECG gating die mean effective dose was 1.8 mSv, and the mean organ dose of heart was 10.3 mGy, this represents an 91% reduction compared with that of pulsing off mode. No differences in quantitative measures of linage quality were found between high pitch cardiac scan mode and others for noise, uniformity, high contrast resolution, linearity.
Conclusion
In patients with a low and stable heart rate, prospectively ECG gated high-pitch cardiac scan provides excellent image quality equivalent to retrospectively ECG gated helical scan-corelated acquisition but reduced effective dose.
목차
Abstract
Ⅰ. 서론
Ⅱ. 대상 및 방법
Ⅲ. 결과
Ⅳ. 고찰
Ⅴ. 결론
참고문헌
키워드
해당간행물 수록 논문
- CT 검사에서 Gantry Aperture의 Isocenter와 피사체 중심의 불 일치 시 선량과 화질의 평가
- CT Guided Interventional Sturdy에서 Conventional Scan Mode 적용에 따른 선량 감소에 관한 연구
- CT실의 환경오염평가와 환경소독에 관한 연구
- Multi-detector Computed Tomography를 이용한 Neck 검사에서 Automatic Exposure Control을 이용한 선량의 최적화
- 폐 소결절 환자를 대상으로 한 CT-guided Wire Marking Procedure에 대한 유용성 평가
- Volume Rendering CT Image의 TACE Reference Data로서의 유용성
- Lumbar Spine Disease 환자의 2D Sagittal Reformation 시 Reformation Angle에 따른 Intervertebral Foramina Image에 대한 고찰
- Adaptive Statistical Iterative Reconstruction 기법의 적용에 따른 Computed Tomography 영상의 화질과 선량 연구
- ECG-gated Cardiac MDCT 검사에서 심장기능 평가를 위한 최적의 수축기말과 확장기말 Phase(%)에 관한 연구
- Multi-detector Computed Tomography에서 환자의 자세 변화에 따른 Esophageal Cancer 진단의 유용성
- Dual Source CT에서 High-pitch Cardiac Scan Mode를 이용한 관상동맥 검사의 유용성
참고문헌
관련논문
의약학 > 방사선과학분야 BEST
- 고준위방사성폐기물 관리와 관련된 사회안전망 구축에 관한 연구
- 민간의료보험 해약 의향 영향 요인: 한국 의료 패널 2019년 2기 자료를 이용하여
- 방사선사 맞춤형 조영제 정맥주사 교육과정 개발
의약학 > 방사선과학분야 NEW
- 재구성 알고리즘과 Butterworth 필터 차단 주파수가 핵의학 영상 화질에 미치는 영향: Prominence Processor 기반 시뮬레이션 연구
- 합성곱 블록 주의 모듈과 U-Net을 활용한 CT 영상 노이즈 제거 기법
- 요오드 조영제 자동 주입 시 외부 가온기 사용이 온도 변화에 미치는 영향
최근 이용한 논문
교보eBook 첫 방문을 환영 합니다!
신규가입 혜택 지급이 완료 되었습니다.
바로 사용 가능한 교보e캐시 1,000원 (유효기간 7일)
지금 바로 교보eBook의 다양한 콘텐츠를 이용해 보세요!