국문 초록
영문 초록
Purpose
To evaluate the feasibility of using relative low tube voltage scan protocols for low-dose about prospective and retrospective ECG gating scan on coronary image quality evaluation with 64-row MDCT.
Materials and methods
A female ART 300 phantom was used to simulate coronary artery of size(5mm in diameter) with four stenosis degrees(0%, 25%, 50% and 75%) at 55bpm heart rate. Cardiac scans were performed on a 64-row MDCT scanner(GE LightSpeed VCT) with rotation time of 350msec, 350mm scan length with proximal-middle coronary artery and two scan methods of prospective scan(axial) and retrospective scan(helical) of 0.2 pitch under each tens different scan protocols. Tube voltage and current were 100kVp/300~600mA, 120kVp/300~600mA, 140kVp/300~400mA. The simulative coronary arteries were filled with contrast media to reach an approximate value 400HU in the lumen. Background noise was measured to describe the basic image quality accordingly. CNR, SNR and contour sharpness represented in slope of CT density curve was calculated as well. Measured stenosis area and rates, described by the percentage area of stenosis on the cross-section images were also calculated. Comparative analysis standard stenosis rate with measured stenosis rate and the error are measured.
Results
Dose(CTDIvol) were 7.4~22.3mGy of prospective scan and 24.3~73.1mGy of retrospective scan by the protocols. Prospective scan method was measured less 30.5% radiation dose than retrospective scan. The corresponding image noise levels described in standard deviation of background signals varied with radiation dose, CNR and SNR mainly varied with tube current. The contour sharpness, which can reflect actual spatial resolution, is affected mainly by tube voltage. When comparing the groups with similar radiation dose of prospective(protocol 10) and retrospective(protocol 1) scans, prospective scan was measured obviously steeper mean slope value than retrospective scan. Significant difference stenosis rate error presented between two groups, which is low dose and high dose. Low dose group was overestimated on stenosis rate. When comparing the groups with similar radiation dose in same scan method protocol, protocols with lower tube voltage gained more accuracy in representing stenosis area and rate.
Conclusion
Prospective scan was estimated similar or higher results than retrospective scan in image quality evaluation as well as lower dose. Dose level and corresponding image quality is relevant to the accuracy of stenosis evaluation on simulated coronary arteries with 64-row MDCT. In this study, we find relative low-dose protocols with acceptable image quality showed a tendency of overestimating stenosis. Furthermore, prospective scan method take an effect reduced dose in condition keep stabile heart rate and using a lower tube voltage and higher tube current to gain accurate imaging result is more applicable than other protocols with the same radiation dose level.
목차
Abstract
Ⅰ. 목적
Ⅱ. 대상 및 방법
Ⅲ. 결과
Ⅳ. 고찰
Ⅴ. 결론
참고문헌
해당간행물 수록 논문
- 관상동맥 CT혈관조영술에서의 운동 인공물에 대한 심전도 보정에 의한 영상재구성 기법
- Dual Energy를 이용한 CT Discography의 평가 방법에 관한 연구
- 다절편 전산화단층촬영을 이용한 관상동맥촬영의 통계적 고찰
- CT검사시 목적장기 이외 부위의 차폐를 통한 피폭선량감소 효과에 관한 연구
- Dual Source CT를 이용한 폐정맥 검사시 전향적 심전도 동조화 기법의 유용성 평가
- 상지의 Direct Injection CT Venogram의 유용성에 대한 고찰
- 관상동맥 CT조영 검사시 선량감소 프로토콜에 관한 연구
- Cardiac CTA의 조영제 주입시간에 관한 연구
- 심부정맥혈전증 검사시 Tracker 기법의 유용성 평가
- 경부 CT검사시 선속경화현상 감소를 위한 환자 자세 및 영상 평가
- Dose Modulation 기법에서 Noise Index 변화에 따른 선량 및 영상 Noise의 변화
- 생체 부분 간이식에서 CT영상을 이용한 간 체적계산법의 Reference Data로서의 유용성
- Small Pulmonary Nodule Localization을 위한 CT Fluoroscopy의 유용성
- 64 Slice MDCT와 심근관류 SPECT, 심초음파를 이용한 심실 기능 지표의 비교
- CT검사시 조영제의 혈관외유출 발생률 감소를 위한 방안 모색
- 급성 뇌경색 환자의 혈전용해제(tPA) 투여를 위한 BEST Protocol의 유용성
- Multi-detector CT장치의 저대조도분해능 평가
- 골밀도 측정에 있어 3D-QCT와 DXA의 유용성 비교
- CT검사로 인한 방사선 피폭이 혈액에 미치는 생물학적 변화에 관한 연구
- 대한전산화단층기술학회지에 게재된 논문의 연구동향
참고문헌
관련논문
의약학 > 방사선과학분야 BEST
- 고준위방사성폐기물 관리와 관련된 사회안전망 구축에 관한 연구
- 민간의료보험 해약 의향 영향 요인: 한국 의료 패널 2019년 2기 자료를 이용하여
- 호주 방사선 의료 전문가 면허제도 및 교육과정 고찰: 진단방사선사를 중심으로
의약학 > 방사선과학분야 NEW
- 소아 복부 CT의 신체 크기 특이적 선량 추정치 평가
- 대한CT영상기술학회지 제27권 제2호 목차
- 소아 CT 데이터의 효과적인 잡음 제거: 성인 CT 데이터 기반 미세조정 기법의 적용
최근 이용한 논문
교보eBook 첫 방문을 환영 합니다!
신규가입 혜택 지급이 완료 되었습니다.
바로 사용 가능한 교보e캐시 1,000원 (유효기간 7일)
지금 바로 교보eBook의 다양한 콘텐츠를 이용해 보세요!