본문 바로가기

추천 검색어

실시간 인기 검색어

학술논문

Effect of 50-Hz Filters on Pattern Electroretinogram

이용수  15

영문명
발행기관
대한안과학회
저자명
Dong Hee Ha Nam Ju Moon Ungsoo Samuel Kim
간행물 정보
『The Korean Journal of Ophthalmology』Vol.39 No.3, 241~245쪽, 전체 5쪽
주제분류
인문학 > 역사학
파일형태
PDF
발행일자
2025.05.31
4,000

구매일시로부터 72시간 이내에 다운로드 가능합니다.
이 학술논문 정보는 (주)교보문고와 각 발행기관 사이에 저작물 이용 계약이 체결된 것으로, 교보문고를 통해 제공되고 있습니다.

1:1 문의
논문 표지

국문 초록

Purpose: Pattern electroretinogram (PERG) is used to evaluate the function of retinal ganglion cells. However, the amplitude of PERG is quite small, making the examination challenging to perform. Waveform noise may be minimized by applying various filters. We aimed to investigate the effect of 50-Hz filters on PERG test results. Methods: This is the retrospective observational study. PERG tests were performed using the RETI-scan system according to the International Society for Clinical Electrophysiology of Vision guidelines. Three types of 50-Hz filters (soft, middle, and hard) were applied. The differences in parameters (N35 peak time, P50 peak time, N95 peak time, P50 amplitude, N95 amplitude, and N95 to P50 ratio) were analyzed. Based on the provided normal range, the changes from normal to abnormal range or vice versa were investigated. Results: A total of 24 waveforms were analyzed. After filtering, the P50 and N95 amplitudes showed a significant reduction of 8% to 15% (P50 amplitude: 5.1 ± 2.7 μV without filter, 4.6 ± 2.3 μV with 50-Hz soft filter, 4.3 ± 2.1 μV with 50-Hz middle filter, 4.3 ± 2.1 μV with 50-Hz hard filter; N95 amplitude: 7.2 ± 4.2 μV without filter, 6.6 ± 3.8 μV with 50-Hz soft filter, 6.3 ± 3.6 μV with 50-Hz middle filter, 6.1 ± 3.6 μV with 50-Hz hard filter). This pattern was more prominent in normal subjects. All latencies except the N35 peak time exhibited no differences between the tests. The N95 to P50 ratio decreased after 50-Hz middle and hard filtering. Considering the normative data, switching between normal and abnormal results was rare. Conclusions: Although peak time was not significantly affected, amplitude was significantly reduced after using 50-Hz filters. Thus, 50-Hz filters can smoothen the waveform. Nevertheless, caution must be exercised while taking readings.

영문 초록

목차

Materials and Methods
Results
Discussion
References

키워드

해당간행물 수록 논문

참고문헌

교보eBook 첫 방문을 환영 합니다!

신규가입 혜택 지급이 완료 되었습니다.

바로 사용 가능한 교보e캐시 1,000원 (유효기간 7일)
지금 바로 교보eBook의 다양한 콘텐츠를 이용해 보세요!

교보e캐시 1,000원
TOP
인용하기
APA

Dong Hee Ha,Nam Ju Moon,Ungsoo Samuel Kim. (2025).Effect of 50-Hz Filters on Pattern Electroretinogram. The Korean Journal of Ophthalmology, 39 (3), 241-245

MLA

Dong Hee Ha,Nam Ju Moon,Ungsoo Samuel Kim. "Effect of 50-Hz Filters on Pattern Electroretinogram." The Korean Journal of Ophthalmology, 39.3(2025): 241-245

결제완료
e캐시 원 결제 계속 하시겠습니까?
교보 e캐시 간편 결제