Purpose: Pattern electroretinogram (PERG) is used to evaluate the function of retinal ganglion cells. However, the amplitude of PERG is quite small, making the examination challenging to perform. Waveform noise may be minimized by applying various filters. We aimed to investigate the effect of 50-Hz filters on PERG test results.
Methods: This is the retrospective observational study. PERG tests were performed using the RETI-scan system according to the International Society for Clinical Electrophysiology of Vision guidelines. Three types of 50-Hz filters (soft, middle, and hard) were applied. The differences in parameters (N35 peak time, P50 peak time, N95 peak time, P50 amplitude, N95 amplitude, and N95 to P50 ratio) were analyzed. Based on the provided normal range, the changes from normal to abnormal range or vice versa were investigated.
Results: A total of 24 waveforms were analyzed. After filtering, the P50 and N95 amplitudes showed a significant reduction of 8% to 15% (P50 amplitude: 5.1 ± 2.7 μV without filter, 4.6 ± 2.3 μV with 50-Hz soft filter, 4.3 ± 2.1 μV with 50-Hz middle filter, 4.3 ± 2.1 μV with 50-Hz hard filter; N95 amplitude: 7.2 ± 4.2 μV without filter, 6.6 ± 3.8 μV with 50-Hz soft filter, 6.3 ± 3.6 μV with 50-Hz middle filter, 6.1 ± 3.6 μV with 50-Hz hard filter). This pattern was more prominent in normal subjects. All latencies except the N35 peak time exhibited no differences between the tests. The N95 to P50 ratio decreased after 50-Hz middle and hard filtering. Considering the normative data, switching between normal and abnormal results was rare.
Conclusions: Although peak time was not significantly affected, amplitude was significantly reduced after using 50-Hz filters. Thus, 50-Hz filters can smoothen the waveform. Nevertheless, caution must be exercised while taking readings.
영문 초록
목차
Materials and Methods
Results
Discussion
References
해당간행물 수록 논문
- Clinical Characteristics and Visual Outcomes of Cat Scratch Disease
- Neopterin Levels and Immune Response in Autoimmune Uveitis in an Experiment
- Pseudoexfoliative Deposits on an Implantable Collamer Lens: A Case Report
- Assessment of Tonometry Methods in Keratoconic Eyes Following Intracorneal Ring Segments Implantation: A Comparative Study
- Effect of 50-Hz Filters on Pattern Electroretinogram
- Reinjection in Patients with Intraocular Inflammation Development after Intravitreal Brolucizumab Injection
- The Efficacy of Sutureless Amnion Membrane Transplantation for Corneal Epithelialization in Delayed Corneal Healing: A Systematic Review
- Delayed-Onset Glaucoma Following Intravitreal Bevacizumab in a Preterm Infant with Retinopathy of Prematurity: A Case Report
- Management Practice for Hordeolum and Chalazion: A Survey of the Korean Society of Ophthalmic Plastic and Reconstructive Surgery (KSOPRS) Members
- Association between Skeletal Muscle Mass and Ocular Perfusion Pressure in Glaucoma
- Treatment of Pterygium on the Refractive Errors: A Systematic Review
참고문헌
교보eBook 첫 방문을 환영 합니다!
신규가입 혜택 지급이 완료 되었습니다.
바로 사용 가능한 교보e캐시 1,000원 (유효기간 7일)
지금 바로 교보eBook의 다양한 콘텐츠를 이용해 보세요!