- 영문명
- An Assessment of the Utility of Respiratory Synchronized Systems in the PET/CT Examination
- 발행기관
- 대한핵의학기술학회
- 저자명
- 성용준(Yong-Jun Seong) 윤석환(Seok-Hwan Yoon) 현준호(Jun-Ho Hyun) 이홍재(Hong-jae Lee) 김진의(Jin-Eui Kim)
- 간행물 정보
- 『핵의학기술』Vol.21 No.1, 34~38쪽, 전체 5쪽
- 주제분류
- 의약학 > 방사선과학
- 파일형태
- 발행일자
- 2017.06.30
4,000원
구매일시로부터 72시간 이내에 다운로드 가능합니다.
이 학술논문 정보는 (주)교보문고와 각 발행기관 사이에 저작물 이용 계약이 체결된 것으로, 교보문고를 통해 제공되고 있습니다.

국문 초록
영문 초록
Purpose During PET/CT examinations, the movements of internal organs caused by respiration are captured in images during multiple breathing cycles, resulting in the increases in tumor size and effects on SUV. Respiratory synchronized systems were used to evaluate tumor sizes and SUV changes. Materials and Methods Biograph mCT 64 was used for the equipment, and RPM and Anzai systems were used for the respiratory synchronized systems. We used point source and micro-phantom for an experimentation. We were performed on 12 patients who had solid tumors discovered at the base of the lung or at the top of the liver from August through September 2016. The PET images of the exhalation-to-breathing state and the CT images of the post-exhalation suspension state were gained to evaluate changes in radioactivity concentration (KBq/mL), SUVmax, cylinder diameter (㎜), and tumor diameter (㎝) under the conventional Static, RPM, and Anzai methods. Results The result of measuring the radioactivity concentration of the point source was RPM 94% and Anzai 91% against Static, respectively. In the two cylinders of different radioactivity in the micro-phantom, the SUVmax increased to RPM 61% and 78%, and Anzai 58% and 77% against Static, whereas the cylinder diameters decreased by RPM -26% and -28%, and Anzai -28% and -26%, each respectively. Among the patients, the SUVmax increased from a minimum of RPM 8.2% to a maximum of 94.4% against Static, and from a minimum of Anzai 7.6% to a maximum of 68.3%, respectively. As for the tumor diameters, a minimum of RPM -7.6% to a maximum of -28.9% were achieved, while the Anzai fell by a minimum of -9.6% to a maximum of -27.7%, respectively. There was no significant difference discovered in the phantom study between the RPM and Anzai, yet there was a meaningful difference in the patients’ tumors (P <0.05). Conclusion The respiratory synchronized systems of RPM and Anzai yielded no significant difference in the phantom study in which the respiration was executed at regular intervals. However, it was discovered that the patients had a meaningful difference for the irregular respiratory cycle and inter-system differences. Still, the respiratory synchronized systems would be useful for the accurate diagnosis and SUV measurement as the tumor decreased in size against the existing Static and the SUV increased.
목차
서론
실험재료 및 방법
결과
고찰
결론
요약
Reference
해당간행물 수록 논문
- 대한핵의학기술학회 투고 논문 분류
- PET/CT 검사에서 Flow mode를 적용한 Respiratory Gating Method 촬영과 추가 Gating 촬영의 비교 및 유용성 평가
- Brain SPECT 검사 시 Dynamic Continuous Mode의 유용성 평가
- Thyroglobulin 검사의 반응조건에 따른 결과 비교 분석
- 방사성 동위원소를 이용한 핵의학과 검사에서 병동 간호사의 방사선 피폭선량 평가
- Thyroglobulin 검사의 반응조건에 따른 결과 비교 분석
- ⁵⁷Co 밀봉선원의 노후도에 따른 균일도의 변동과 중첩사용에 관한 유용성 평가
- 갑상선 섭취율 측정에서 선원의 길이와 부피 증가에 따른 영향
- 물리적 보정된 핵의학 영상 평가
- 설문조사를 바탕으로 부서 간 만족도 향상을 위한 개선 안
- 전신 뼈 스캔의 오류 감소를 위한 프로세스 구축과 적용 효과
- PET-CT 검사 시 호흡 동조 시스템들의 유용성 평가
- Westgard Multi-Rules의 효율적 적용과 조치사항의 개선
참고문헌
관련논문
의약학 > 방사선과학분야 BEST
- 고준위방사성폐기물 관리와 관련된 사회안전망 구축에 관한 연구
- 민간의료보험 해약 의향 영향 요인: 한국 의료 패널 2019년 2기 자료를 이용하여
- 호주 방사선 의료 전문가 면허제도 및 교육과정 고찰: 진단방사선사를 중심으로
의약학 > 방사선과학분야 NEW
- 소아 복부 CT의 신체 크기 특이적 선량 추정치 평가
- 대한CT영상기술학회지 제27권 제2호 목차
- 소아 CT 데이터의 효과적인 잡음 제거: 성인 CT 데이터 기반 미세조정 기법의 적용
최근 이용한 논문
교보eBook 첫 방문을 환영 합니다!
신규가입 혜택 지급이 완료 되었습니다.
바로 사용 가능한 교보e캐시 1,000원 (유효기간 7일)
지금 바로 교보eBook의 다양한 콘텐츠를 이용해 보세요!
