The Usability Assessment of the Self-Produced Table for the Quality Control of Mobile CT
이용수 0
- 영문명
- The Usability Assessment of the Self-Produced Table for the Quality Control of Mobile CT
- 발행기관
- 대한CT영상기술학회
- 저자명
- Dong Jun Jang Seong Min Choi Dan Bi Lee
- 간행물 정보
- 『대한CT영상기술학회지』대한CT영상기술학회지 제24권 제1호, 1~8쪽, 전체 8쪽
- 주제분류
- 의약학 > 방사선과학
- 파일형태
- 발행일자
- 2022.03.31
4,000원
구매일시로부터 72시간 이내에 다운로드 가능합니다.
이 학술논문 정보는 (주)교보문고와 각 발행기관 사이에 저작물 이용 계약이 체결된 것으로, 교보문고를 통해 제공되고 있습니다.

국문 초록
영문 초록
Since there are no tables customized for mobile CT examinations, it is difficult to place the AAPM phantom to the isocenter during the quality control. This produced a phantom table for the quality control of Mobile CT, assessing its utility through conducting a quantitative evaluation through CNR value comparison between with and without the use of the table on low contrast resolution images among 20 phantom images evaluation, along with a qualitative assessment conducted on the scale of 1 to 5 by 5 quality control operators.
The obtained average CNR values from low contrast resolution images in the order of the table being used and not used are -0.01069 ± 0.00149 vs -0.01485 ± 0.00208, -0.01277 ± 0.00131 vs -0.01526 ± 0.00215, -0.01419 ± 0.00144 vs -0.01671 ± 0.00277, -0.01651 ± 0.00140 vs -0.01836 ± 0.00199, -0.01542 ± 0.0188 vs -0.01836 ± 0.00199 from A-row, -0.01132 ± 0.00134 vs -0.01300 ± 0.00222, -0.1202 ± 0.00129 vs -0.01369 ± 0.00145, -0.001391 ± 0.00163 vs -0.01460 ± 0.00139, -0.01607 ± 0.00185 vs -0.01683 ± 0.00182, -0.01502 ± 0.00170 vs -0.01656 ± 0.00125 from B-row, indicating higher absolute values of CNR when the table is being used than not being used, which corresponds to more suitable images. Furthermore, the average score of all items in qualitative assessment has increased from 3.64 to 4.24, suggesting statistically significant margin of increase especially on the items of efficiency and the quality of images. In conclusion, placing the phantom on a more accurate isocenter through the exploitation of the table produced through this research provides phantom images with higher quality along with the decrease of unnecessary efforts from the operators.
목차
I. INTRODUCTION
II. MATERIAL AND METHODS
III. RESULT
IV. DISCUSSION
V. CONCLUSIONS
REFERENCES
해당간행물 수록 논문
- The Usability Assessment of the Self-Produced Table for the Quality Control of Mobile CT
- Effectiveness of Tube Voltage Selection of 70 kV in Abdominal Dual Energy CT Protocol
- Effect of 3D Printing Material and Thickness on CT Image Quality and Shielding
- Comparison of Accuracy of Coronary Artery Calcium Score between Artificial Intelligence and Manual MethodComparison of Accuracy of Coronary Artery Calcium Score between Artificial Intelligence and Manual Method
- A Study on the Correlation of Left Atrial Volume Change according to Left Ventricular Contraction in Atrial Fibrillation
- 대한CT영상기술학회지 제24권 제1호 목차
참고문헌
관련논문
의약학 > 방사선과학분야 BEST
- 고준위방사성폐기물 관리와 관련된 사회안전망 구축에 관한 연구
- 민간의료보험 해약 의향 영향 요인: 한국 의료 패널 2019년 2기 자료를 이용하여
- 호주 방사선 의료 전문가 면허제도 및 교육과정 고찰: 진단방사선사를 중심으로
의약학 > 방사선과학분야 NEW
- 무증상 치밀 유방 여성에서 인공지능 기반 컴퓨터 보조 진단 장치의 유용성 연구
- Cardiofocal Collimator를 이용한 SPECT 검사 활용 증대 방안에 대한 연구: Brain SPECT를 중심으로
- 심장초음파의 좌심실이완기능장애와 대사증후군의 상관관계 분석
최근 이용한 논문
교보eBook 첫 방문을 환영 합니다!
신규가입 혜택 지급이 완료 되었습니다.
바로 사용 가능한 교보e캐시 1,000원 (유효기간 7일)
지금 바로 교보eBook의 다양한 콘텐츠를 이용해 보세요!