This study was purpose to evaluation of Modulation Transfer Function in Measurements by using the International electrotechnical commission standard(IEC 62220-1) which were edge device each angle by using edge method. In this study was Aero(Konica, Japan) image receptor which is a indirect Flat panel detector(FPD) was used. The size of matrix 1994 × 2430 (14″× 17″inch) which performed 12 bit processing and pixel pitch is 175 μm. The results of shown as MTF measurements at IEC standard. The amount of data seemed reasonable and at an MTF value of 0.1 the spatial frequencies were 2.56 cycles/mm at an angle of 2.4°. MTF value of 0.5 the spatial frequencies were 1.32 cycles/mm at an angle of 2.4°. This study were to evaluate MTF by setting each angle 2.0°∼2.8° degrees the most effective optimal edge angle and to suggest the quantitative methods of measuring by using IEC.
목차
Ⅰ. 서 론
Ⅱ. 대상 및 방법
Ⅲ. 결 과
Ⅳ. 고 찰
Ⅴ. 결 론
REFERENCES
키워드
해당간행물 수록 논문
참고문헌
관련논문
의약학 > 방사선과학분야 BEST
- 고준위방사성폐기물 관리와 관련된 사회안전망 구축에 관한 연구
- 민간의료보험 해약 의향 영향 요인: 한국 의료 패널 2019년 2기 자료를 이용하여
- 호주 방사선 의료 전문가 면허제도 및 교육과정 고찰: 진단방사선사를 중심으로
의약학 > 방사선과학분야 NEW
- 소아 복부 CT의 신체 크기 특이적 선량 추정치 평가
- 대한CT영상기술학회지 제27권 제2호 목차
- 소아 CT 데이터의 효과적인 잡음 제거: 성인 CT 데이터 기반 미세조정 기법의 적용
최근 이용한 논문
교보eBook 첫 방문을 환영 합니다!
신규가입 혜택 지급이 완료 되었습니다.
바로 사용 가능한 교보e캐시 1,000원 (유효기간 7일)
지금 바로 교보eBook의 다양한 콘텐츠를 이용해 보세요!