구획모델에 기반하는 폐의 기능적 잔기용량 계측기법

이용수  5

영문명
Measurement of Functional Residual Capacity Based on Compartmental Model
발행기관
한국산업기술융합학회(구. 산업기술교육훈련학회)
저자명
김군진(Goon-Jin Kim) 최성수(Sung-Su Chio) 김경아(Kyung-Ah Kim) 차은종(Eun-Jong Cha)
간행물 정보
『산업기술연구논문지』산업기술교육훈련논문지 제15권 2호, 1~12쪽, 전체 12쪽
주제분류
공학 > 공학일반
파일형태
PDF
발행일자
2010.06.30
4,240

구매일시로부터 72시간 이내에 다운로드 가능합니다.
이 학술논문 정보는 (주)교보문고와 각 발행기관 사이에 저작물 이용 계약이 체결된 것으로, 교보문고를 통해 제공되고 있습니다.

1:1 문의
논문 표지

국문 초록

영문 초록

The present study confirmed through experiment the superiority of new functional residual capacity (FRC) measuring techniques using a compartmental model to the existing technique in measurement characteristics. Assuming through deductive inference from a known compartmental model that the inspiration and‐expiration process is a constant air flow and the initial N2 concentration in the lung at the moment of measuring by N2 washout is known, we took the pulmonary ventilation system as the compartmental model. Because the compartmental model is a mathematical model explaining the variation of N2 ratio(FE) at the end of expiration with given FRC size, initial N2 concentration of FRC, and average air flow, reversely we can obtain FRC from the model if the initial N2 concentration of FRC and the average air flow are given and the variation of FE can be quantified. In order to test the validity of these measuring methods, we should have experimented with a subject, but because the subject’s absolute FRC could not be measured, we used syringe apparatus similar to a subject’s breathing only through the mouth and showing exact internal capacity(FRC). Experiment was conducted with the internal capacity of the syringe set at 0.1[L], 0.6[L], 1.1[L], 1.6[L], 2.1[L], and 2.6[L], respectively. The internal capacity of syringe was measured by the methods using FE and average air flow, which are necessary for the new methods, and by the existing method using expiratory air capacity and the N2 concentration in the expiratory air bag. When the result of estimating the internal capacity of syringe was analyzed in each experiment, the correlation efficient between the reference value(initially set FRC) and the estimated value was 0.9629 in the existing method, and 0.9956 in the new method. The standard deviation of absolute error was ±0.262[L] in the existing method, and ±0.087[L] in the new methods above. The present study confirmed through syringe experiment that, in N2 washout, the new FRC measuring techniques are more accurate than the existing method. It is considered necessary to test the validity of the measuring techniques devised in this study through additional clinical experiments.

목차

Ⅰ. 서 론
Ⅱ. 기능적 잔기용량 추정이론
Ⅲ. 실 험
Ⅳ. 자료 분석
Ⅳ. 고찰 및 결론

키워드

해당간행물 수록 논문

참고문헌

교보eBook 첫 방문을 환영 합니다!

신규가입 혜택 지급이 완료 되었습니다.

바로 사용 가능한 교보e캐시 1,000원 (유효기간 7일)
지금 바로 교보eBook의 다양한 콘텐츠를 이용해 보세요!

교보e캐시 1,000원
TOP
인용하기
APA

김군진(Goon-Jin Kim),최성수(Sung-Su Chio),김경아(Kyung-Ah Kim),차은종(Eun-Jong Cha). (2010).구획모델에 기반하는 폐의 기능적 잔기용량 계측기법. 산업기술연구논문지, 15 (2), 1-12

MLA

김군진(Goon-Jin Kim),최성수(Sung-Su Chio),김경아(Kyung-Ah Kim),차은종(Eun-Jong Cha). "구획모델에 기반하는 폐의 기능적 잔기용량 계측기법." 산업기술연구논문지, 15.2(2010): 1-12

결제완료
e캐시 원 결제 계속 하시겠습니까?
교보 e캐시 간편 결제