- 영문명
- Evaluation of QC Value Variation and Overlapping Use According to Aging of ⁵⁷Co Sealed Source
- 발행기관
- 대한핵의학기술학회
- 저자명
- 이종훈(Jong-Hun Lee) 조성길(Sung-kil Cho) 심동오(Dong-Oh Shim)
- 간행물 정보
- 『핵의학기술』Vol.21 No.1, 50~53쪽, 전체 4쪽
- 주제분류
- 의약학 > 방사선과학
- 파일형태
- 발행일자
- 2017.06.30

국문 초록
영문 초록
Purpose The change in uniformity due to the decreasement in dose should be known. ⁵⁷Co sealed sources are easy to manage and use as QC sources replacing 99mTc. Overlapping sealed sources are expected to show variations in dose due to attenuation between sealed sources. Materials and Methods A total of three experiments were conducted. The first experiment is to observe the change in the degree of senescence of the ⁵⁷Co sealed source. The second experiment is to compare the single source and overlaped source at similar doses. In the third experiment, the sources of different doses were compared on each other to determine the changes due to the attenuation between the overlapping sources. Results The results of the first experiment did not exceed the acceptable range. but each crew showed a difference. There was no statistically significant difference in the measurement of uniformity on second and third experiment Conclusion It is believed that a ⁵⁷Co sealed source can be used as a superimposed source. It is not only economical but also convenient to use. daily uniformity measurements will help reduce scan time and speed up the testing process
목차
서론
실험재료 및 방법
결과
결론
고찰
요약
Referances
해당간행물 수록 논문
참고문헌
최근 이용한 논문