- 영문명
- Development of Bladder Phantom and Image Evaluation of Lesion in the Vicinity according to Filling and Empty Bladder
- 발행기관
- 대한핵의학기술학회
- 저자명
- 박찬록(Chan-Rok Park) 김재일(Jae-Il Kim) 이홍재(Hong-Jae Lee) 김진의(Jin-Eui Kim)
- 간행물 정보
- 『핵의학기술』Vol.20 No.1, 24~27쪽, 전체 4쪽
- 주제분류
- 의약학 > 방사선과학
- 파일형태
- 발행일자
- 2016.05.30

국문 초록
영문 초록
Purpose In this study, we evaluated image quality of lesion in the vicinity according filling and empty bladder developing bladder phantom. Materials and Methods Bladder phantom was developed by modifying NEMA IEC body phantom. Air-balloon was described as bladder and 6 insert were set as lesion in the vicinity bladder according to distance. The images were evaluated in terms of % BV, comparison of SUV and peak count rate (Single, Random and True count). Results % BV was decreased far away from bladder. There were different for SUV about 7.8 ± 3.8 % between filling and empty bladder. True count rate was decreased about 38 %. Single and random count were increased about 44, 61%. Conclusion When the lesion is close to bladder, noise is increased. That’s why prior to PET-CT scan, It is important to urinate. and It helps patient to get the accurate result.
목차
서론
실험재료 및 방법
결과
결론 및 고찰
요약
REFERENCES
해당간행물 수록 논문
참고문헌
최근 이용한 논문