- 영문명
- Comparative Evaluation of 25-OH-VitD and 25-OH-VitD3 by Radioimmunoassay
- 발행기관
- 대한핵의학기술학회
- 저자명
- 이영지(Young-ji Lee) 박지혜(Ji-hye Park) 임수연(Soo-yeon Lim) 천준홍(Jun-hong Cheon) 이선호(Sun-ho Lee)
- 간행물 정보
- 『핵의학기술』Vol.24 No.1, 33~38쪽, 전체 6쪽
- 주제분류
- 의약학 > 방사선과학
- 파일형태
- 발행일자
- 2020.09.30

국문 초록
영문 초록
Purpose Vitamin D is essential for maintaining bone health, controling cell proliferation or differentiation, strengthening immune function by controlling calcium metabolism in the body. Vitamin D deficiency can lead to increase the risk of rickets, osteoporosis, cardiovascular disease, diabetes and cancer. Especially, South Korea is one of the highest population proportion of vitamin D deficiency. Accurate determination of levels of 25-OH-VitD or 25-OH-VitD3 in blood serum is required for the diagnosis and treatment of vitamin D deficiency. In this study, radioimmunoassay of 25-OH-VitD and 25-OH-VitD3 was performed and compared to evaluate the effectiveness of Vitamin D radioimmunoassay.
Materials and Methods Serum 25-OH-VitD and 25-OH-VitD3 levels were measured using radioimmunoassay. The interrelationship, reproducibility and population distribution rate were evaluated. In addition, the internal quality control was performed at Asan Medical Center from April 2017 to June 2019 and the result of external quality control (Interagency proficiency evaluation) of first and second half of 2018 hosted by the Korean Society of Nuclear Medicine Technology (KSNMT). Both tests were measured by same manufacturer's reagent.
Results 25-OH-VitD showed a strong positive correlation on 97 samples, as 25-OH-VitD3 x 0.9 + 0.3 (R >0.9). In repeated measurement, the average Diff(%) value of the reproducibility evaluation of 25-OH-VitD and 25-OH-VitD3 were 7.7% and 7.4%, respectively. Population distribution results showed no statistically significant differences(p>0.05). The resultant value of internal quality control, which measured from April, 2017 to June 2019 in Blood test room of Nuclear Medicine at Asan Medical Center, showed average (CV%) 6.2% and 6.8%, respectively. As a result of the external quality control (interagency proficiency evaluation) Z value obtained under 2.0, as shown below;
Conclusion The interrelationship, reproducibility, population distribution rate, internal quality control and external quality control between 25-OH-VitD and 25-OH-VitD3 radioimmunoassay shows superior outcome. Radioimmunoassay, which can be alone measured in the blood as 25-OH-VitD or 25-OH-VitD3, is considered suitable screening tests for the diagnosis of vitamin D deficiency.