- Performance Evaluation of a Pinhole Collimator According to the Aperture Diameter
- 발행기관
- 대한핵의학기술학회
- 저자명
- 이광훈(Kwang Hun Lee) 안병호(Byung Ho An) 김수영(Soo Young Kim) 최성욱(Sung Wook Choi)
- 간행물 정보
- 『핵의학기술』Vol.18 No.1, 104~109쪽, 전체 6쪽
- 주제분류
- 의약학 > 방사선과학
- 파일형태
- 발행일자
- 2014.05.30

국문 초록
영문 초록
Purpose: Conventional pinhole scintigraphy offers a high resolution and generally use for thyroid scan or bone scintigraphy. the aim of this study is to evaluate performance of each aperture according to the diameter size. Materials and Methods: In this study 2 mm, 4 mm, 6 mm, 8 mm diameter pinhole collimators were mounted on Siemens E.CAM systems. In order to evaluate performance evaluation of each aperture, we acquired projection image by using SPECT for evaluating FWHM, resolution test and static image for evaluating thyroid phantom test. Results: As a result of FWHM showed 2.2 mm FWHM, 3.2 mm FWHM, 5.4 mm FWHM, 7.5 mm FWHM per each aperture in the resolution test. SNR, uniformity, contrast were acquired result from thyroid phantom test comparing general image and delayed image. as a result of SNR showed 6.55, 8.47, 6.2, 5.23 in case of general image and 5.25, 5.01, 5.38, 5.82 in case of delayed image per each aperture. as a result of uniformity showed 0.152, 0.118, 0.161, 0.19 in case of general image and 0.19, 0.199, 0.185, 0.171 in case of delayed image per each aperture. as a result of contrast showed 1.31, 1.19, 1.15, 1.01 in case of general image and 1.09, 1.08, 1.04, 1 in case of delayed image per each aperture. Conclusion: Resolution and sensitivity were heavily influenced by aperture size. and also we found advantage of delayed image as a result of thyroid phantom test.
목차
서론
실험재료 및 방법
결과
결론
요약
REFERENCES
해당간행물 수록 논문
참고문헌
최근 이용한 논문
교보eBook 첫 방문을 환영 합니다!
신규가입 혜택 지급이 완료 되었습니다.
바로 사용 가능한 교보e캐시 1,000원 (유효기간 7일)
지금 바로 교보eBook의 다양한 콘텐츠를 이용해 보세요!