- The Evaluation of SUV Variations According to the Errors of Entering Parameters in the PET-CT Examinations
- 발행기관
- 대한핵의학기술학회
- 저자명
- 김지아(Jia Kim) 홍건철(Gun Chul Hong) 이혁(Hyeok Lee) 최성욱(Seong Wook Choi)
- 간행물 정보
- 『핵의학기술』Vol.18 No.1, 43~48쪽, 전체 6쪽
- 주제분류
- 의약학 > 방사선과학
- 파일형태
- 발행일자
- 2014.05.30

국문 초록
영문 초록
Purpose: In the PET/CT images, The SUV (standardized uptake value) enables the quantitative assessment according to the biological changes of organs as the index of distinction whether lesion is malignant or not. Therefore, It is too important to enter parameters correctly that affect to the SUV. The purpose of this study is to evaluate an allowable error range of SUV as measuring the difference of results according to input errors of Activity, Weight, uptake Time among the parameters. Materials and Methods: Three inserts, Hot, Teflon and Air, were situated in the 1994 NEMA Phantom. Phantom was filled with 27.3 MBq/mL of 18F-FDG. The ratio of hotspot area activity to background area activity was regulated as 4:1. After scanning, Image was re-reconstructed after incurring input errors in Activity, Weight, uptake Time parameters as ±5%, 10%, 15%, 30%, 50% from original data. ROIs (region of interests) were set one in the each insert areas and four in the background areas. SUVmean and percentage differences were calculated and compared in each areas. Results: SUVmean of Hot. Teflon, Air and BKG (Background) areas of original images were 4.5, 0.02. 0.1 and 1.0. The min and max value of SUVmean according to change of Activity error were 3.0 and 9.0 in Hot, 0.01 and 0.04 in Teflon, 0.1 and 0.3 in Air, 0.6 and 2.0 in BKG areas. And percentage differences were equally from -33% to 100%. In case of Weight error showed SUVmean as 2.2 and 6.7 in Hot, 0.01 and 0.03 in Tefron, 0.09 and 0.28 in Air, 0.5 and 1.5 in BKG areas. And percentage differences were equally from -50% to 50% except Teflon area’s percentage deference that was from -50% to 52%. In case of uptake Time error showed SUVmean as 3.8 and 5.3 in Hot, 0.01 and 0.02 in Teflon, 0.1 and 0.2 in Air, 0.8 and 1.2 in BKG areas. And percentage differences were equally from 17% to -14% in Hot and BKG areas. Teflon area's percentage difference was from -50% to 52% and Air area's one was from -12% to 20%. Conclusion: As shown in the results, It was applied within ±5% of Activity and Weight errors if the allowable error range was configured within 5%. So, The calibration of dose calibrator and weighing machine has to conduct within ±5% error range because they can affect to Activity and Weight rates. In case of Time error, it showed separate error ranges according to the type of inserts. It showed within 5% error when Hot and BKG areas error were within ±15%. So we have to consider each time errors if we use more than two clocks included scanner’s one during the examinations.
목차
서론
실험재료 및 방법
결과
결론
고찰
요약
REFERENCES
해당간행물 수록 논문
참고문헌
최근 이용한 논문
교보eBook 첫 방문을 환영 합니다!
신규가입 혜택 지급이 완료 되었습니다.
바로 사용 가능한 교보e캐시 1,000원 (유효기간 7일)
지금 바로 교보eBook의 다양한 콘텐츠를 이용해 보세요!