- Correlation of SUV According to the Body Weight and Body Surface Area in PET/CT Imaging
- 발행기관
- 대한핵의학기술학회
- 저자명
- 박주선(Ju-Seon Park) 도용호(Yong-Ho Do) 김정수(Jung-Soo Kim) 이홍재(Hong-Jae Lee) 김진의(Jin-Eui Kim)
- 간행물 정보
- 『핵의학기술』Vol.17 No.2, 44~47쪽, 전체 4쪽
- 주제분류
- 의약학 > 방사선과학
- 파일형태
- 발행일자
- 2013.11.30

국문 초록
영문 초록
Purpose: There are several methods of measuring SUV in PET/CT imaging including SUVbw which uses the body weight, SUVbsa which that uses body surface area and SUVlbm which uses lean body mass. Currently, Seoul National University Hospital uses SUVbw method which minimizes the variability. In this study, we compared and analyzed the correlation between SUVbw and SUVbsa according to patients' body mass index. Materials and Methods: Using Biograph mCT40 (Siemens, Germany), we conducted ¹⁸F-FDG PET/CT imaging on 70 patients (41 males, 29 females; ages 58.04±12.44). We classified the patients as underweight (BMI<20), normal weight (20≦BMI<25), overweight (25≦BMI<30), obese (30≦BMI<35) and severely obese (35≦BMI) according to the patient's sex, age and BIM. Then, bone, liver and lungs were set as ROI for calculation of maximum values of SUVbw and SUVbsa, through Syngo.via VA11A analysis program. Results: Comparing the five groups divided according to the BMI by the standard differences between SUVbw to SUVbsa, SUVmax was measured to be 0.66±0.15, 0.78±0.35, 0.77±0.21, 1.00±0.44, 1.53±0.38 for bones in underweight, normal weight, overweight, obese and severely obese groups, respectively. For liver, values of SUVmax were 1.64±0.16, 2.06±0.34, 2.19±0.21, 2.52±0.21 and 2.74±0.40 in the same order. And for lung, values of SUVmax were 0.69±0.33, 0.54±0.17, 0.62±0.23, 0.83±0.29, 1.03±0.30. Conclusion: By comparing and analyzing the differences between SUVbw and SUVbsa in this study, it was found that the differences between SUVbw and SUVbsa increased as patient's BMI increased. Thus, there is room for error in the values of SUV depending on the methods of calculations, and appropriate methods must be applied according to the circumstances in clinical settings.
목차
서론
실험재료 및 방법
결과
결론
요약
REFERENCES
해당간행물 수록 논문
참고문헌
최근 이용한 논문
교보eBook 첫 방문을 환영 합니다!
신규가입 혜택 지급이 완료 되었습니다.
바로 사용 가능한 교보e캐시 1,000원 (유효기간 7일)
지금 바로 교보eBook의 다양한 콘텐츠를 이용해 보세요!