- 영문명
- Comprehension and Appropriate Use of a Flood Table on a Gamma Camera
- 발행기관
- 대한핵의학기술학회
- 저자명
- 김재일(Jae Il Kim) 임정진(Jeong Jin Im) 김진의(Jin Eui Kim) 김현주(Hyun Joo Kim)
- 간행물 정보
- 『핵의학기술』Vol.15 No.1, 29~33쪽, 전체 5쪽
- 주제분류
- 의약학 > 방사선과학
- 파일형태
- 발행일자
- 2011.04.30

국문 초록
영문 초록
Background and Purpose: Uniformity is the one of the important quality control features with respect to gamma cameras. To maintain adequate uniformity, we must acquire suitable flood table (= flood map) data because the flood table effects energy, and the type or dose of input radiation. Therefore, in this study we evaluated the difference in uniformity when uniformity does not match between the type of input radiation and the flood table data or collimator type. Subjects and Methods: For input radiation, we prepared 370 MBq of ⁵⁷Co, ⁹⁹ᵐTc, and 201Tl. Using SKYLight (Philips) and Infinia gamma cameras (GE), we acquired nine uniformity data that were corrected by technetium, cobalt flood table and did not corrected image for the three sources. Additionally, we acquired two uniformity images with a collimator that were corrected by intrinsic and extrinsic flood tables. Using this data, we evaluated and compared the uniformity values. Results: In the case of the SKYLight gamma camera, the uniformities of the images that matched between the input radiation and flood table with respect to ⁹⁹ᵐTc and ⁵⁷Co were better than the unmatched uniformity (3.96% vs. 5.69% ; 4.9% vs. 5.91%). However, because there was no thallium flood table, the uniformities of images at Tl were significantly incorrect (7.49%, 7.03%). The uniformities of the Infinia gamma camera had the same pattern as the SKYLight gamma camera (3.7% vs. 4.5%). Moreover, the uniformity of the ⁹⁹ᵐTc image acquired with a collimator and corrected by an extrinsic flood table was better than the intrinsic flood table (3.96% vs. 6.28%). Conclusion: Correcting an image by a suitable flood table can help achieve better uniformity for a gamma camera. Therefore, we have to acquire images with suitable uniformity correction, and update the flood table periodically. Whenever we acquire a nuclear medicine image, we always have to check the appropriate flood table according to the acquired condition.
목차
서론
재료 및 방법
결과
결론
요약
REFERENCES
해당간행물 수록 논문
참고문헌
최근 이용한 논문
교보eBook 첫 방문을 환영 합니다!
신규가입 혜택 지급이 완료 되었습니다.
바로 사용 가능한 교보e캐시 1,000원 (유효기간 7일)
지금 바로 교보eBook의 다양한 콘텐츠를 이용해 보세요!