국문 초록
영문 초록
Purpose: The aim of study is to find accuracy of pocket dosimeter in measuring exposed dose in compared with survey meter and to compare exposed dose according as Nuclear medicine exams. Materials and Method: First, radiation dose to point source(185 MBq,370 MBq, ..., 1665 MBq, 1850 MBq) were measured in using a pocket dosimeter and a survey meter. Second, radiation dose to 12 patients injected ¹⁸F-FDG 370 MBq were measured in using a pocket dosimeter and a survey meter. Third, radiation dose to 10 patients injected ⁹⁹ᵐTc-DPD 925 MBq were measured in using a pocket dosimeter and a surveymeter. Result: The average is 70.12±39.36 μSv/h in measurement of point source with Surveymeter and 5±3.06 μSv in measurement of point source with Pocket dosimeter. The average is 25.04±6.16 μSv/h in measurement of PET/CT patients with Surveymeter and 2.41±0.51 μSv in measurement of PET/CT with Pocket dosimeter. The average is 8.58±0.96 μSv/h in measurement of Bone Scan patients with Surveymeter and 1 μSv in measurement of Bone Scan patients with Pocket dosimeter. Significant difference found between Survey meter value and Pocket dosimeter value in all experimentation (p<0.001). Conclusion: Accoring to rusult Wearing Pocket dosimeter is usefulnee in manerage of exposed dose in nucler medicine exams.
목차
서론
실험재료 및 방법
결과
고찰
결론
요약
REFERENCE
해당간행물 수록 논문
참고문헌
최근 이용한 논문
교보eBook 첫 방문을 환영 합니다!
신규가입 혜택 지급이 완료 되었습니다.
바로 사용 가능한 교보e캐시 1,000원 (유효기간 7일)
지금 바로 교보eBook의 다양한 콘텐츠를 이용해 보세요!