The purpose of this study is to evaluate the clinical risk of spinal radiosurgery by calculating the dose difference due to dose calculation algorithm and multi-leaf collimator positioning error. The images acquired by the CT simulator were recalculated by correcting the multi-leaf collimator position in the dose verification program created using MATLAB and applying stoichiometric calibration and Monte Carlo algorithm. With multi-leaf collimator positioning error, the clinical target volume (CTV) showed a dose difference of up to 13% in the dose delivered to the 95% volume, while the gross tumor volume (GTV) showed a dose difference of 9%. The average dose delivered to the total volume showed dose variation from -8.9% to 9% and -10.1% to 10.2% for GTV and CTV, respectively. The maximum dose delivered to the total volume of the spinal cord showed a dose difference from -14.2% to 19.6%, and the dose delivered to the 0.35 ㎤ volume showed a dose difference from -15.5% to 19.4%. In future research, automating the linkage between treatment planning systems and dose verification programs would be useful for spinal radiosurgery.
목차
Ⅰ. 서 론
Ⅱ. 대상 및 방법
Ⅲ. 결 과
Ⅳ. 고 찰
Ⅴ. 결 론
REFERENCES
키워드
해당간행물 수록 논문
참고문헌
최근 이용한 논문
교보eBook 첫 방문을 환영 합니다!
신규가입 혜택 지급이 완료 되었습니다.
바로 사용 가능한 교보e캐시 1,000원 (유효기간 7일)
지금 바로 교보eBook의 다양한 콘텐츠를 이용해 보세요!