- 영문명
- Assessment of Dose and Image Quality according to the Change of Distance from Source to Image Receptor and the Examination Posture during the Skull Lateral Radiography
- 발행기관
- 대한방사선과학회(구 대한방사선기술학회)
- 저자명
- 김은혜(Eun-Hye Kim) 주영철(Young-Cheol Joo) 김한용(Han-Yong Kim) 김동환(Dong-Hwan Kim)
- 간행물 정보
- 『방사선기술과학』제45권 제6호, 483~489쪽, 전체 7쪽
- 주제분류
- 의약학 > 방사선과학
- 파일형태
- 발행일자
- 2022.12.31

국문 초록
영문 초록
This study proposes a new skull lateral examination, and provides an improved examination environment for patients and radiologists. The study was divided into three groups. One group was divided into the SID (source to image receptor distance) 110 ㎝ and 180 ㎝ in the skull lateral posture, the other group The other group was divided into an position in contact with the detector and an position without contact with the detector, and the other group was divided into male and female groups, considering that the difference in shoulder width between adult males and females would affect the dose and image quality. For dose evaluation, the ESD (entrance surface dose) was measured at the EAM (external auditory meatus), and the conditions were applied equally at 70 ㎸p, 200 ㎃, and 10 ㎃s. For image quality evaluation, SNR (signal to noise ratio) and CNR (contrast to noise ratio) were measured in frontal sinus, EAM, and sella turcica. As a result of ESD comparison, when sid 110 ㎝ to sid 180 ㎝ was changed among the three groups, ESD values decreased the most to 729.18±4.62 μ㏉ and 224.18±0.74 μ㏉ at 180 ㎝ (p<0.01). The values of SNR and CNR were statistically significant (p<0.01), but there was no qualitative difference. This shows that when the SID is 180 ㎝, it is possible to reduce the dose without lowering the image quality. So, It is suggested that the SID 180 ㎝ is used without contacting the detector when examining the skull lateral.
목차
Ⅰ. 서 론
Ⅱ. 대상 및 방법
Ⅲ. 결 과
Ⅳ. 고 찰
Ⅴ. 결 론
REFERENCES
키워드
해당간행물 수록 논문
참고문헌
최근 이용한 논문
교보eBook 첫 방문을 환영 합니다!
신규가입 혜택 지급이 완료 되었습니다.
바로 사용 가능한 교보e캐시 1,000원 (유효기간 7일)
지금 바로 교보eBook의 다양한 콘텐츠를 이용해 보세요!