학술논문
Giant Retinal Tear by Jamming of 23-gauge Instrument in the Cannula during Vitrectomy for Dense Vitreous Hemorrhage
이용수 0
- 영문명
- 발행기관
- 대한검안학회
- 저자명
- Sun Jung Eum Jae Pil Shin
- 간행물 정보
- 『Annals of optometry and contact lens』Vol.14 No.4, 223~226쪽, 전체 4쪽
- 주제분류
- 의약학 > 기타의약학
- 파일형태
- 발행일자
- 2015.12.30
국문 초록
영문 초록
Purpose:We report a rare case of intraoperative giant retinal tear caused by completely jammed 23-gauge instrument in the cannula during vitrectomy for dense vitreous hemorrhage.
Case summary: A 41-year-old man presented with visual disturbance in the left eye. Uncorrected visual acuity was 20/20 for right eye and hand motion in the left eye which was not corrected. The patient was diagnosed with dense vitreous hemorrhage related to branch retinal vein occlusion in the left eye. We performed 23-gauge vitrectomy, and the vitreous cutter was firmly jammed inside the cannula intraoperatively. The jammed cannula had to be removed along with the cutter and a new cannula was reinserted through the sclerotomy site. A giant retinal break was detected around the replaced cannula during peripheral vitrectomy. Perfluorocarbon liquid was used to unfold the inverted flap of the retinal tear and the operation was finished by applying laser photocoagulation and silicone oil tamponade.
Conclusions: The 23-gauge vitrectomy instrument may be jammed in the cannula during vitrectomy for dense vitreous hemorrhage, and the trapped vitreous can cause retinal break by traction. Therefore, dense vitreous hemorrhage around the cannula should be carefully removed to prevent iatrogenic intraoperative retinal break prior to core vitrectomy.
목차
CASE REPORT
DISCUSSION
REFERENCES
해당간행물 수록 논문
참고문헌
최근 이용한 논문
교보eBook 첫 방문을 환영 합니다!
신규가입 혜택 지급이 완료 되었습니다.
바로 사용 가능한 교보e캐시 1,000원 (유효기간 7일)
지금 바로 교보eBook의 다양한 콘텐츠를 이용해 보세요!