- 영문명
- A Case of Rapidly Occurring Intraoperative Clouding of Hydrophilic Multifocal Intraocular Lens
- 발행기관
- 대한검안학회
- 저자명
- 박연꽃(Yeon Ggoch Park) 최규룡(Kyu Ryong Choi)
- 간행물 정보
- 『Annals of optometry and contact lens』Vol.14 No.3, 163~165쪽, 전체 3쪽
- 주제분류
- 의약학 > 기타의약학
- 파일형태
- 발행일자
- 2015.09.30

국문 초록
영문 초록
Purpose: To report a case of intraoperative clouding of hydrophilic multifocal intraocular lens (Lentis Mplus?? LS-313, Oculentis GmbH., Berlin, Germany) during cataract surgery.
Case summary: A 69-year-old woman underwent intraocular lens exchange in her left eye due to clouding of an intraocular lens occurred intraoperatively. This phenomenon was caused by the rapid change in temperature of a very cold IOL being implanted into the eye.
Conclusions: It is important to prevent a rapid temperature change during intraocular lens implantation. Particularly in the country that have winter season and fluctuating temperatures, careful caution will be necessary.