- The Effect of MLC Leaf Motion Constraints on Plan Quality and Delivery Accuracy in VMAT
- 발행기관
- 대한방사선과학회(구 대한방사선기술학회)
- 저자명
- 김연래(Kim Yon-Lae) 정진범(Chung Jin-Beom) 이정우(Lee Jeong-woo) 신영주(Shin Young-Joo) 강동진(Kang Dong-Jin) 정재용(Jung Jae-Yong)
- 간행물 정보
- 『방사선기술과학』방사선기술과학 제42권 제3호, 217~222쪽, 전체 6쪽
- 주제분류
- 의약학 > 방사선과학
- 파일형태
- 발행일자
- 2019.06.30

국문 초록
영문 초록
The purpose of this study is to evaluate the dose distribution by gantry rotation and MLC moving speed on
treatment planning system(TPS) and linear accelerator. The dose analyzer phantom(Delta 4) was scanned by CT simulator for treatment planning. The planning target volumes(PTVs) of prostate and pancreas was prescribed 6,500 cGy, 5,000 cGy on VMAT(Volumetric Modulated Arc Therapy) by TPS while MLC speed changed. The analyzer phantom was irradiated linear accelerator using by planned parameters. Dose distribution of PTVs were evaluated by the homogeneity index, conformity index, dose volume histogram of organ at risk(rectum, bladder, spinal cord, kidney). And irradiated dose analysis were evaluated dose distribution and conformity by gamma index. The PTV dose of pancreas was 4,993 cGy during 0.1㎝/deg leaf and gantry that was the most closest prescribed dose(5,000 cGy). The dose of spinal cord, left kidney, and right kidney were accessed the lowest during 0.1 ㎝/deg, 1.5 ㎝/deg, 0.3 ㎝/deg. The PTV dose of prostate was 6,466 cGy during 0.1 ㎝/deg leaf and gantry that was the most closest prescribed dose(6,500 cGy). The dose of bladder and rectum were accessed the lowest during 0.3 ㎝/deg, 2.0 ㎝/deg. For gamma index, pancreas and prostate were analyzed the lowest error 100% at 0.8, 1.0 ㎝/deg and 99.6% at 0.3, 0.5 ㎝/deg. We should used the optimal leaf speed according to the gantry rotation if the treatment cases are performed VMAT.
목차
Ⅰ. 서 론
Ⅱ. 대상 및 방법
Ⅲ. 결 과
Ⅳ. 고 찰
Ⅴ. 결 론
해당간행물 수록 논문
참고문헌
최근 이용한 논문
교보eBook 첫 방문을 환영 합니다!
신규가입 혜택 지급이 완료 되었습니다.
바로 사용 가능한 교보e캐시 1,000원 (유효기간 7일)
지금 바로 교보eBook의 다양한 콘텐츠를 이용해 보세요!