In this study, we have researched the effectiveness of silicone pad. A distribution of scatter ray in mammography
was evaluated using Monte-Carlo (MC) simulation technique and then a silicone pad was applied to remove the
scatter ray for improving image quality. Molybdenum target and Molybdenum filter combination made a difference of
59.8% to a number of photon at 17.5 keV. On the other hand, Tungsten target and Rhodium filter showed a variation of
24.5% at 20 keV. Mean 68 of SNR was increased in Selenia and mean 1.04 of SNR was raised in Senographe. Silicone
pad was significantly effective to reduce the scatter ray that was generated by primary X-ray. It can decrease an absorption rate of scatter ray to patient body and whilst it improve the image quality from increasing SNR.