학술논문
The Comparison of Central and Mean True-Net Power (Pentacam) in Calculating IOL-Power After Refractive Surgery
이용수 0
- 영문명
- 발행기관
- 대한안과학회
- 저자명
- Jeong-Ho Yi Joo Youn Shin Byoung Jin Ha Sang Woo Kim Beom Jin Cho Eung Kweon Kim Tae-im Kim
- 간행물 정보
- 『The Korean Journal of Ophthalmology』Vol.23 No.1, 1~5쪽, 전체 5쪽
- 주제분류
- 의약학 > 의학일반
- 파일형태
- 발행일자
- 2009.03.30
국문 초록
영문 초록
Purpose: To compare the accuracy of central true net corneal power (cTNP) and mean true net corneal power
(mTNP) of the Pentacam system to give a keratometry (K) reading for calculating IOL (intraocular lens) power in
eyes following refractive surgery.
Methods: Refraction, an automated K-reading (Km), cTNP and mTNP were measured for 15 eyes that required
cataract surgery and had previously undergone refractive surgery. The difference between postoperative manifest
refraction values and target refraction values calculated with the SRK/T formula using cTNP were compared with
the one using mTNP.
Results: The mean deviation from the desired post-cataract refractive outcome was 0.60 diopter (D) ±0.47
(standard deviation) using cTNP; 0.75±0.54 using mTNP (p=0.386). The actual refraction was within ±0.50D of the
intended refraction for 60% (cTNP) and 33.3% (mTNP) of eyes, and within ±1.00D for 93% (cTNP) and 66.7%
(mTNP) of eyes.
Conclusions: Although not statistically significant, the cTNP showed better accuracy than mTNP to give a keratometry
(K) reading for post-refractive surgery eyes requiring cataract surgery.
Korean J Ophthalmol 2009;23:1-5 ⓒ 2009 by the Korean Ophthalmological Society.