- 영문명
- Evaluation of Radiation Doses of Dental Portable Equipment
- 발행기관
- 대한방사선과학회(구 대한방사선기술학회)
- 저자명
- 박훈희(Park Hoon-Hee) 강병삼(Kang Byung-Sam)
- 간행물 정보
- 『방사선기술과학』방사선기술과학 제41권 제5호, 445~450쪽, 전체 6쪽
- 주제분류
- 의약학 > 방사선과학
- 파일형태
- 발행일자
- 2018.10.30

국문 초록
영문 초록
We aim to evaluate safety of radiation by measuring leakage dose and patient(phantom) incident dose of ZEN-PX II dental portable equipment developed by G company. Measurement for leakage dose of equipment is conducted on the top, at the bottom, on the left, on the right and at the back. Dose measurement incident on the subject with the area dosimeter when using the phantom and measurement the leakage dose of equipment when using the phantom are evaluated. Comparing the right with the highest leakage dose as a 0 cm, 25 cm, 50 cm, 75 cm and 100 cm dose measurement at the measurement height of 100 cm, 64.2 uR was reduced to 47.3 uR in the senser mode 0.32sec. Even in film mode it was measured at 414.4 uR and about 27% lower at 162.6 uR. As the result of this study, when the irradiation time is 2 sec the right side dose is 290.5 uR and sensor mode is 0.32 sec the right side dose is 64.2 uR
목차
Ⅰ. 서 론
Ⅱ. 대상 및 방법
Ⅲ. 결 과
Ⅳ. 고 찰
Ⅴ. 결 론