- 영문명
- Comparison of Estimated and Measured Doses of Dual-energy Computed Tomography
- 발행기관
- 대한방사선과학회(구 대한방사선기술학회)
- 저자명
- 김영균(Kim Yung-Kyoon) 김연민(Kim Yon-Min)
- 간행물 정보
- 『방사선기술과학』방사선기술과학 제41권 제5호, 405~411쪽, 전체 7쪽
- 주제분류
- 의약학 > 방사선과학
- 파일형태
- 발행일자
- 2018.10.30

국문 초록
영문 초록
We will provide basic data on the evaluation of patient dose in terms of DECT quality control by comparing the equipment-provided dose with the measured dose according to the configuration method of the X-ray generator by the manufacturer of the dual-energy CT unit. For computed tomography (CT) equipment, Discovery 750HD, Aquilion ONE GENESIS Edition, and Somatom Definition Flash were used. The CTDI vol value was measured by inserting the Unfors Xi ion chamber into a 32 cm PMMA acryl Phantom.
The results of estimated CTDI vol DECT and measured CTDI vol showed that the dose difference between DECT 80 + 140kVp of G company was at least 0.51% and -1.90% max, and measured CTDI vol was slightly lower (p<0.05). The difference of 80 + 140 kVp of S company was the minimum of 5.84% and the maximum of 7.52% (p<0.05). The measured CTDI vol was less than estimated CTDI vol . The C company’s 80 + 135 kVp showed a difference of at least 7.58% and a maximum of 13.58% (P<0.05), and all of measured CTDI vol was less. The linearity of exposure dose for all DECT equipment was very linearly reflected with R 2 being 0.97 or above, and the measured dose of the ionization chamber was less than the predicted dose of the monitor.
목차
Ⅰ. 서 론
Ⅱ. 대상 및 방법
Ⅲ. 결 과
Ⅳ. 고 찰
Ⅴ. 결 론
키워드
해당간행물 수록 논문
참고문헌
최근 이용한 논문
교보eBook 첫 방문을 환영 합니다!
신규가입 혜택 지급이 완료 되었습니다.
바로 사용 가능한 교보e캐시 1,000원 (유효기간 7일)
지금 바로 교보eBook의 다양한 콘텐츠를 이용해 보세요!