- 영문명
- Low Dose CT Virtual Colonography: Comparison Study
- 발행기관
- 대한CT영상기술학회
- 저자명
- 김정훈(Jeong hoon Kim) 정재현(Je Hyeon Jeong) 김기홍(Ki Hong Kim)
- 간행물 정보
- 『대한CT영상기술학회지』대한전산화단층기술학회지 제10권 제1호, 77~81쪽, 전체 5쪽
- 주제분류
- 의약학 > 방사선과학
- 파일형태
- 발행일자
- 2008.04.30

국문 초록
영문 초록
Purpose
We compared our study to evaluate an usefulness of a low dose CT virtual colonography.
Materials and methods
During February 2006, 20 patients with colon cancer were taken CT virtual colonography. A 16 MDCT (Somatom Sensation 16, Siemens, Germany) was used for this study. The variable mA according to the dose modulation program with care dose 4D was used in supine position, and the constant mA with lower dose than that was used in prone position. 5 experts including 1 radiologist, 2 radiological technologist charged in 3D processing, and 2 radiological technologist charged in CT evaluated the dose and images.
Results
The value of CTDIv in using care dose 4D was a mean 9.00mGy, and constant mA was a mean 3.90mGy. There was not different between the images including shape, definition, and flexibility of variable mA and constant mA (p< 0.05).
Conlclusion
Our data suggest that the low dose CT virtual colonography is useful in the aspect of dose and image.