- 영문명
- A study of appropriate expose condition in Chest CT for Care Dose 4D
- 발행기관
- 대한CT영상기술학회
- 저자명
- 서재영(Jae Yeong Seo)
- 간행물 정보
- 『대한CT영상기술학회지』대한전산화단층기술학회지 제10권 제1호, 123~128쪽, 전체 6쪽
- 주제분류
- 의약학 > 방사선과학
- 파일형태
- 발행일자
- 2008.04.30

국문 초록
영문 초록
Purpose
We intended to inspect the optimum exposure condition which lessens the quality deterioration of picture at the chest examination using Care Dale 4D
Materials and methods
We measured the Volume, Noise, CTDIvol, and DIP of the models according to the changes of kVp and mAs, by inserting nine models, manufactured randomly, into the phantom of the chest in manikin. We used 64 MDCT and Wizard Workstation for examination equipment and analyzed the relationship comparatively using SPSS(Ver. 12) for statistical analysis
Results
The Volume of the model according to the change of kVp and Quality reference mAs showed attentive value(P>0.05), remaining similar. On the contrary, the value of Noise indicated the decrease (P<0.05) according to the increase of kVp and Quality reference mAs and CTDIvol and DLP increased according to the increase of kVp and Quality reference mAs.
Conclusion
For the chest CT usfug Care Dale 4D, we can minimize the exposure dose of the patient by using proper kVp and Quality reference mAs without quality deterioration of picture.