- 영문명
- Investigation about dose efficiency of MDCT
- 발행기관
- 대한CT영상기술학회
- 저자명
- 신상보(Sang Bo Shin) 조원홍(Won Hong Cho) 조남수(Nam Soo Cho)
- 간행물 정보
- 『대한CT영상기술학회지』대한전산화단층기술학회지 제7권 제1호, 201~208쪽, 전체 8쪽
- 주제분류
- 의약학 > 방사선과학
- 파일형태
- 발행일자
- 2005.03.30

국문 초록
영문 초록
This study was found out change factors of dose efficiency in multi- detector CT. The dose efficiency was analyzed by change of technical scan parameter such as kVp, mA, pitch, SFOV, beam collimation with 8 & 16 channels CT.
The results obtained were as follows;
1. Dose efficiency was higher 20 mm beam collimation than 10 mm beam collimation in all devices.
2. Dose efficiency was higher at small focal spot than large focal spot in all devices.
3. There was no change of dose efficiency about change of pitch in all devices.
4. There was no change of dose efficiency about change of SFOV in all devices.
5. The 16 channels MDCT was higher dose efficiency than 8 channels MDCT.