Cardiac CT검사에 있어서 심박동수 및 그 변화가 Image Quality 및 Optimized Reconstruction Window Phase에 미치는 영향에 대한 평가
이용수 17
- 영문명
- Evaluation of Influence on Image quality and Optimized reconstruction window by Heart rate and it’s Variation on Cardiac CT study.
- 발행기관
- 대한CT영상기술학회
- 저자명
- 김상현(Sang Hyun Kim) 김영옥(Young Ok Kim) 대창민(Chang Min Dae) 민관홍(Kwan Hong Min) 김성룡(Swig Yong Kim)
- 간행물 정보
- 『대한CT영상기술학회지』대한전산화단층기술학회지 제7권 제1호, 184~193쪽, 전체 10쪽
- 주제분류
- 의약학 > 방사선과학
- 파일형태
- 발행일자
- 2005.03.30

국문 초록
영문 초록
Purpose
To investigate the influence on image quality of coronary CT angiography by heart rate and its variation.
Material and Methods
We studied 119 patients who had undergone coronary CT angiography using a heart rate responsive gating algorithm and adaptive multi-cycle reconstructions. Ranges of heart rate(bpm: beats per minutes) before the scan were classified as follows: 60 bpm < (n= 28), 61-70 bpm(n=35), 71-80 bpm (n=26), 81-90 bpm(n=l6), 91-100 bpm(n=9), 100 bpm > (n=5). Two radiologists rated the image quality as follows; 4: excellent, 3: good, 2: fair, 1: bad. Influence on image quality by heart rate and its variation during the scan was analyzed.
Results
Variations of heart rate were showed as follows: 0-10 bpm < (n=86 patients), 11-20 bpm(n=26), 20 > bpm(n=7). High variation of heart rate showed more poor image quality(0-10 bpm: 3.550.62, 11-20 bpm: 3.120.52, 20 > bpm: 2.120.64). Image quality was decreased as heart rate was higher(60< bpm: 3.570.7, 61-70 bpm: 3.260.87, 71-80 bpm: 3.460.8, 81-90 bpm: 3.040.64, 91-100 bpm: 3.22 0.97, 100> bpm: 3.000.71), but image quality was more affected by the variation of the heart rate than by the higher heart rate. The selection of an optimized reconstruction window is mainly affected by the heart rate, not by its variation. There is tendency for systolic phase(50%) to have a superior image quality than diastolic phase(75%) at higher heart rates(above 70 bpm) regardless of its variation.
Conclusion
We suggest that the variation of heart rate during the scan is important factor on image quality of coronary CT angiography. But it does not affect to the selection of an optimized reconstruction window.