- 영문명
- Evaluation of the Utility of a Volumetric Modulated Arc Therapy Specific Patient Quality Assurance using Software-based Quality Assurance System
- 발행기관
- 대한방사선과학회(구 대한방사선기술학회)
- 저자명
- 강동진(Dong-Jin Kang) 정재용(Jae-Yong Jung) 신영주(Young-Joo Shin) 민정환(Jung-Whan Min) 김연래(Yon-Lae Kim) 권경태(Kyung-Tae Kwon)
- 간행물 정보
- 『방사선기술과학』방사선기술과학 제41권 제1호, 1~8쪽, 전체 8쪽
- 주제분류
- 의약학 > 방사선과학
- 파일형태
- 발행일자
- 2018.02.28

국문 초록
영문 초록
The purpose of this study is to evaluate the usefulness of a software-based quality assurance system based on Volumetric Modulated Arc Therapy treatment plan. Evaluate treatment plan through the D VH analysis, PTV mean dose ( ) and PTV 95% dose( ) compare the MFX based on original treatment plan, Average error rate was 0.9±0.6%, 1.0±0.8%, respectively. Measuring point dose using phantom and ion chamber, the average error rate between the ionization chamber and MFX was 0.9±0.7%, 1.1±0.7% (high dose region), 1.1±0.9%, 1.2±0.7% (low dose region). The average gamma though of MFX and Delta 4PT is 98.7±1.2%, 98.4±1.3%, respectively. Through this study, A software based QA system that simplifies hardware based QA procedures that involve a lot of time and effort. It can be used as a simple and useful tool in clinical practice.