- During the Chest PA, the Consideration of Patients’ Effective Dose
- 발행기관
- 대한영상의학기술학회
- 저자명
- 박상만(Sang Man Park) 이상전(Sang Jeon Lee) 안미섭(Mi Seop An) 문희석(Hee Seok Moon) 조남수(Nam Soo Cho)
- 간행물 정보
- 『대한영상의학기술학회 논문지』2009년 춘계학술대회 논문지, 123~130쪽, 전체 8쪽
- 주제분류
- 의약학 > 방사선과학
- 파일형태
- 발행일자
- 2009.03.30

국문 초록
영문 초록
Purpose : When examining chest PA, to know whole body x-ray damage level exactly by measuring patients’ effective dose.
Materials and methods : The experiment equipment is REVOLUTION XQI chest private x-ray machine of GE Company, Glass dosimeter (DOSE ACE [Asahi Technograju Company]) is used for measuring expose dose. The experimentway is to expose with the conditions FFD 180cm, 100kVp(HVL 4.08mmAl), 3.08mAs, Field Size 43cm*43cm, calculating air absorbing dose by inserting Glass dosimeter organ position endowed organ load coefficient named by ICRP inside ART(ANDERSON RADIATION THERAPHY) Phantom To indicate organ absorbing dose by air absorbing dose in each position multiplies organ and air mass suction coefficient, and then to know Effective Dose by multiplying each organ load coefficient. We found 1 expose effective dose by averaging total 10 times expose, and knew errors 5 times with doing the same process.
Results : Examining the above, the average ESD was about 0.1116mGy (SD 4.89, CV 0.044), effective dose was about 29.8420Sv.
Conclusion : During the chest PA, the effective dose of patients damaged is about 0.0298mSv, 5 times repetition result is SD 1.04, CV 0.03489.
목차
Abstact
Ⅰ. 서론
Ⅱ. 대상 및 방법
Ⅲ. 결과
Ⅳ. 고찰 및 결론
참고문헌
해당간행물 수록 논문
참고문헌
최근 이용한 논문
교보eBook 첫 방문을 환영 합니다!
신규가입 혜택 지급이 완료 되었습니다.
바로 사용 가능한 교보e캐시 1,000원 (유효기간 7일)
지금 바로 교보eBook의 다양한 콘텐츠를 이용해 보세요!