- 영문명
- Quality Control Manual & Research on the Actual Condition about the Telos Stress Device
- 발행기관
- 대한영상의학기술학회
- 저자명
- 임종천(Jong Cheon Lim) 길은심(Shim Kil Eun) 김치헌(Chie Heon Kim) 김재용(Jae Yong Kim)
- 간행물 정보
- 『대한영상의학기술학회 논문지』2009년 춘계학술대회 논문지, 107~112쪽, 전체 6쪽
- 주제분류
- 의약학 > 방사선과학
- 파일형태
- 발행일자
- 2009.03.30

국문 초록
영문 초록
Purpose : Telos Strees Device is medical instrument that be known whether joint ligament is ruptured or not. However the control manual & objective standard is not in Korea. This study will make an alternative plan.
Materials and methods : we set a standard value in research with phone, and next visit from hospital to hospital for actual survey to 14 object Telos Stress Device.
Telos use the cushion as a matter of feature of equipment, but in this study be used accessorry equipment by onself for actual survey. Telos device 5 times executes pressure prices 5,10,15,20,15 daN respectively and produces a standard value. This time, change applies 1daN with 40.98KG and the error range does with ±1daN.
Results:l4 Telos Stress Device application results the result came out Over average group is 3 &Normal verage group is 8 & Under average group is 2.
One was caused by in the fault in comliance with an exteranl shock and the management did not become accomplished from equiqment 42.9%.
목차
Abstact
Ⅰ. 서론
Ⅱ. 대상 및 방법
Ⅲ. 결과
Ⅳ. 고찰 및 결론
참고문헌