- 영문명
- STABILITY OF MAXILLARY POSITION AFTER LEFORT I OSTEOTOMY USING BIODEGRADABLE PLATES AND SCREWS
- 발행기관
- 대한구강악안면외과학회
- 저자명
- 김봉철(Bong-Chul Kim) 정영수(Young-Soo Jung)
- 간행물 정보
- 『대한구강악안면외과학회지』대한구강악안면외과학회지 제33권 제5호, 499~503쪽, 전체 5쪽
- 주제분류
- 의약학 > 내과학
- 파일형태
- 발행일자
- 2007.10.30

국문 초록
영문 초록
Introduction: In orthognathic surgery, internal fixation has been usually done with titanium plates and screws. Recently, Biodegradable plates and screws have been frequently used but the reports of long term results of postoperative stability are rare, especially after maxillary reposition in orthognathic surgery.
Objective: In order to clarify the clinical utility of self-reinforced bioresorbable poly-70L / 30DL-Lactide miniplates & screws in maxillary fixation after LeFort I osteotomy, this study examined the postsurgical changes in maxilla and complications of biodegradable plates and screws.
Study design: Nineteen patients who had undergone maxillary internal fixation using biodegradable plates and screws were evaluated radiographically and clinically. A comparison study of the changes in maxilla position after surgery in all 19 patients was performed with 1-week, 1-month, 3-months, 6-months and/or 1-year postoperative lateral cephalograms by tracing. Complication of the biodegradable plates and screws was evaluated by
follow-up roentgenograms and clinical observation. And one-way ANOVA test was used for statistical analysis.
Results: The position of the maxillary bone was stable after surgery and was not changed significantly from 1 week to 1 year after operation. And we could not find any complication of biodegradable plates and screws.
Conclusions: Internal fixation of the maxilla after LeFort I osteotomy using self-reinforced biodegradable plates and screws is a reliable method for maintaining postoperative position of the maxilla after LeFort I osteotomy.
목차
해당간행물 수록 논문
참고문헌
최근 이용한 논문
교보eBook 첫 방문을 환영 합니다!
신규가입 혜택 지급이 완료 되었습니다.
바로 사용 가능한 교보e캐시 1,000원 (유효기간 7일)
지금 바로 교보eBook의 다양한 콘텐츠를 이용해 보세요!