- 영문명
- Comparison of the Refractive Results Measured by Ultrasound and Partial Coherence Interferometers
- 발행기관
- 대한안과학회
- 저자명
- 신정아 정성근,Jeing Ah Shin, MD, Sung Kun Chung, MD, PhD
- 간행물 정보
- 『대한안과학회지』Ophthalmological Society,volume54,number5, 723~727쪽, 전체 5쪽
- 주제분류
- 인문학 > 역사학
- 파일형태
- 발행일자
- 2013.05.15

국문 초록
영문 초록
Purpose: To compare ocular biometry measured by applanation ultrasonography and IOL Master®, and evaluate the accuracies of the refractive outcome after cataract surgery. Methods: The biometries of 76 cataractous eyes were measured using ultrasonography and IOL Master®. The SRK-T formula was employed to predict the patient’s implanted IOL power. Two months after cataract surgery, the refractive outcome was determined, and results from the 2 different biometry methods were compared. Results: There were no statistically significant differences in axial length (AXL) and anterior chamber depth (ACD) between ultrasonography and IOL Master® (p = 0.501). When using ultrasonography, the mean absolute error (MAE) of the ultrasonography and IOL Master® was 0.53 ± 0.30 diopter (D), 0.55 ± 0.41 D, respectively. The difference between the 2 biometry methods was not statistically significant (p = 0.110). Conclusions: Although the difference was not statistically significant, AXL measured by IOL Master® was longer and ACD measured by IOL Master® was deeper than when measured by A-scan. This difference was more pronounced in patients with a short AXL. The accuracy of IOL power calculation was similar between the 2 devices.
목차
해당간행물 수록 논문
참고문헌
최근 이용한 논문
교보eBook 첫 방문을 환영 합니다!
신규가입 혜택 지급이 완료 되었습니다.
바로 사용 가능한 교보e캐시 1,000원 (유효기간 7일)
지금 바로 교보eBook의 다양한 콘텐츠를 이용해 보세요!