- 영문명
- Comparison of Corneal Curvatures Obtained with Different Devices
- 발행기관
- 대한안과학회
- 저자명
- 김소윤 정성근,Soyoon Kim, MD, Sung Kun Chung, MD, PhD
- 간행물 정보
- 『대한안과학회지』Ophthalmological Society,volume53,number5, 618~625쪽, 전체 8쪽
- 주제분류
- 의약학 > 의학일반
- 파일형태
- 발행일자
- 2012.05.15
국문 초록
영문 초록
Purpose: To evaluate device efficacy using the corneal curvature value. Methods: Prospectively, 35 patients (70 eyes) were enrolled in the present study. Three sets of corneal curvature values were obtained using the autorefractor (RK-F1Ⓡ), manual keratometer (OM-2Ⓡ), partial coherence interferometry keratometer (IOL MasterⓇ), wavefront analyzer (KR-1WⓇ), and videokeratography (Orbscan IIⓇ). Repeatability of each device was evaluated by coefficient of variation, standard deviation, and intraclass correlation coefficient. RM-ANOVA on ranks was used to compare the differences in corneal curvatures among the devices. The Bland-Altman plot was performed to assess measurement agreement among the devices. Results: The coefficient of variation values from each device ranged from 2.92% (IOL masterⓇ) to 3.06% (Orbscan IIⓇ), and the values of intraclass correlation coefficient ranged from 0.965 (KR-1WⓇ) to 0.997 (IOL masterⓇ). Compared with the manual keratometer, there was a maximum corneal curvature difference of 1.23 D in KR-1WⓇ, while the other devices had differences less than 0.82 D. Conclusions: When assessing corneal curvature, the repeatability values were similar among the 5 devices, although a difference greater than 1 D was observed when comparing the KR-1WⓇ to the manual keratometer. J Korean Ophthalmol Soc 2012;53(5):618-625
목차
해당간행물 수록 논문
참고문헌
최근 이용한 논문
교보eBook 첫 방문을 환영 합니다!
신규가입 혜택 지급이 완료 되었습니다.
바로 사용 가능한 교보e캐시 1,000원 (유효기간 7일)
지금 바로 교보eBook의 다양한 콘텐츠를 이용해 보세요!