- Comparison of Aberrations in Korean Normal Eyes Measured With Two Different Aberrometers
- 발행기관
- 대한안과학회
- 저자명
- 염정훈 최석규 김진형 이도형,Jung Hoon Yum, MD, Suk Kyue Choi, MD, Jin Hyoung Kim, MD, Do Hyung Lee, MD, PhD
- 간행물 정보
- 『대한안과학회지』Ophthalmological Society,volume50,number12, 1789~1794쪽, 전체 6쪽
- 주제분류
- 의약학 > 기타의약학
- 파일형태
- 발행일자
- 2009.12.15

국문 초록
영문 초록
Purpose: To compare ocular higher order aberrations measured by two different aberrometers in a sample of normal eyes. Methods: We included 41 normal eyes of Koreans in this study. Ocular aberration data were obtained through three measurements per eye using Zywave and OPD-Scan devices. Spherical equivalent and higher order aberrations calculated in the central 6 mm zone and expressed as root mean square (RMS) values were analyzed. Results: A comparison of measurements between the Zywave and OPD-Scan devices demonstrated no statistically significant differences in the RMS values of total higher order aberration (p=0.11), but significant differences were detected in the RMS values of total spherical aberration, total coma and total trefoil (p<0.01). Conclusions: The two different aberrometers that we tested are suitable for taking repeated measurements and are internally consistent, but not interchangeable. J Korean Ophthalmol Soc 2009;50(12):1789-1794