- 영문명
- Comparison of Pentacam with Orbscan
- 발행기관
- 대한안과학회
- 저자명
- 신영주 김남희 김대환,Young Joo Shin, M.D., Nam Hee Kim, M.D., Dae Hwan Kim, M.D.
- 간행물 정보
- 『대한안과학회지』Ophthalmological Society,volume48,number5, 637~641쪽, 전체 5쪽
- 주제분류
- 인문학 > 역사학
- 파일형태
- 발행일자
- 2007.05.31

국문 초록
영문 초록
Purpose: To compare the corneal measurements of orbscan and pentacam in normal subjects. Methods: From April 10 th 2006 to May 10 th 2006, subjects were tested by automated refractokeratometry (Canon RK-1, Japan), Pentacam (OCULUS, Germany), and Orbscan (Orbtek Inc, USA). Measurements included central, superior, nasal, inferior and temporal corneal thickness, radius of anterior elevation and posterior elevation, keratometric value and axis of the 4 mm zone, and anterior chamber depth. Results: A total of 96 eyes from 48 patients were included, with 13 men and 35 women. Average age was 25.4 years old. In comparing the Orbscan and Pentacam, central, superior and temporal corneal thicknesses of the Pentacam were significantly thicker than those of the Orbscan (paired t-test, p=.000, p=.000, p=.003). The mean K value of the 4mm zone of the Orbscan was steeper than that of the Pentacam (paired t-test, p=.000). The anterior chamber depth of the Pentacam was deeper than that of the Orbscan (paired t-test, p=.000), but all of the Pentacam measurements were strongly correlated to those of the Orbscan. Conclusions: The measurements of the Pentacam and the Orbscan were significantly different, but they were strongly correlated. Both machines can be useful in clinical applications with partial compensation.
목차
해당간행물 수록 논문
참고문헌
최근 이용한 논문
교보eBook 첫 방문을 환영 합니다!
신규가입 혜택 지급이 완료 되었습니다.
바로 사용 가능한 교보e캐시 1,000원 (유효기간 7일)
지금 바로 교보eBook의 다양한 콘텐츠를 이용해 보세요!