- Homogeneous Dose Planning to Paranasal Sinus with the Partial Attenuation filters and Wedged Beams in 6 MV Photon Beam
- 발행기관
- 대한방사선종양학회
- 저자명
- 최태진 (Tae Jin Choi) 이호준 (Ho Joon Lee) 김옥배 (Ok Bae Kim)
- 간행물 정보
- 『대한방사선종양학회지』제11권 제1호, 183~191쪽, 전체 9쪽
- 주제분류
- 의약학 > 종양학
- 파일형태
- 발행일자
- 1993.06.30

국문 초록
영문 초록
The homogeneous dose planning is one of the most important roles in radiation therapy. But, it is not easy to obtain a homogeneous dose to paranasal sinus region including the ethmoidal sinus with conventional irradiation techniques.
In this experimental study, the authors tried to get a homogeneous dose at PNS region, but the nasal cartirage does not exceed the tolerance dose. with anterior-posterior beam and two both lateral wedged beams. Used three fields were shielded with full thickness of blocks to preserve the eye-balls and with blocks of one half value layer to create a homogeneous dose at the whole treatment volume.
The dose computations are based on the three dimensonal structure with modified scatter contributions of partial shielders and attenuated beams in 6 MV photon beams.
The dose dirtributions of mid-plane is examined with Kodak verification films and teflon-embedded TLD rod (1 mm diameter and 6 mm length) to confirm the computed dose.
In our study, the whole PNS regions have shown within 85% of the resultant isodose curves with relatively homogeneous dose distribution.
The results of dose computation and measurements are agree well within 5% uncertainties.