- 영문명
- Measurement of Intrarater Reliability and Interrater Reliability of a Portable Digital Pressure Algometer
- 발행기관
- 한국전문물리치료학회
- 저자명
- 이원휘(Won-hwee Lee) 오재섭(Jae-seop Oh) 박규남(Kyue-nam Park) 권오윤(Oh-yun Kwon) 송미혜(Mi-hye Song) 박호동(Ho-dong Park) 이경중(Kyoung-joung Lee)
- 간행물 정보
- 『한국전문물리치료학회지』한국전문물리치료학회지 제15권 제3호, 62~69쪽, 전체 8쪽
- 주제분류
- 의약학 > 의학일반
- 파일형태
- 발행일자
- 2008.09.30
국문 초록
영문 초록
The purpose of this study was to measure intrarater and interrater reliability of a portable digital pressure algometer. Fifty healthy subjects were recruited for this study. Pressure pain thresholds of splenius capitis, levator scapular, and upper trapezius muscles were measured using by FPK algometer and portable digital pressure algometer. Three trials were done on each subject by two examiners. Intraclass correlation coefficients (ICCs) were used to determine the reliability of each measure. The intrarater reliability of the FPK algometer was good and excellent (ICC .63~.79) and the interrater reliability was poor and good (ICC .15~.57). The intrarater reliability of a portable digital pressure algometer was excellent (ICC .75~.86) and the interrater reliability was poor and good (ICC .35~.61). This result suggests that the intrarater reliability and interrater reliability of a portable digital pressure algometer were better than an FPK algometer.
목차
Abstract
Ⅰ. 서론
Ⅱ. 연구방법
Ⅲ. 결과
Ⅳ. 고찰
Ⅴ. 결론
인용문헌
Ⅰ. 서론
Ⅱ. 연구방법
Ⅲ. 결과
Ⅳ. 고찰
Ⅴ. 결론
인용문헌
해당간행물 수록 논문
참고문헌
최근 이용한 논문
교보eBook 첫 방문을 환영 합니다!
신규가입 혜택 지급이 완료 되었습니다.
바로 사용 가능한 교보e캐시 1,000원 (유효기간 7일)
지금 바로 교보eBook의 다양한 콘텐츠를 이용해 보세요!