- 영문명
- A Study to Establish Target Exposure Index for Chest Radiography
- 발행기관
- 대한방사선과학회(구 대한방사선기술학회)
- 저자명
- 정회원(Hoi-Woun Jeong) 민정환(Jung-Whan Min)
- 간행물 정보
- 『방사선기술과학』제47권 제3호, 167~173쪽, 전체 7쪽
- 주제분류
- 의약학 > 방사선과학
- 파일형태
- 발행일자
- 2024.06.30
국문 초록
영문 초록
This study purpose to establish an appropriate target exposure index(EIT) using dose area product(DAP) and exposure index(EI) based on chest radiography. First, the system response experiment was conducted with radiation quality of RQA5 to compare the dosimetry and dose area product of equipment. Next, EI and DAP were acquired and analyzed while varying the dose in the diagnostic at 70 kVp using a human body model phantom. The signal to noise ratio(SNR) of the obtained results was analyzed in the diagnostic with in the diagnostic reference level(DRL) application range. The DRL at percentage 25% had a dose of 0.17 mGy and EI was 83, and at percentage 75% the dose was 0.68 mGy and EI was 344. As the dose increased, the SNR in the subdiaphragm increased. To set the EIT, calibration must first be performed using a dosimeter and set within the DRL range to reflect the needs of the medical institution.
목차
Ⅰ. 서 론
Ⅱ. 실험기기 및 방법
Ⅲ. 결 과
Ⅳ. 고 찰
Ⅴ. 결 론
REFERENCES
키워드
해당간행물 수록 논문
참고문헌
최근 이용한 논문
교보eBook 첫 방문을 환영 합니다!
신규가입 혜택 지급이 완료 되었습니다.
바로 사용 가능한 교보e캐시 1,000원 (유효기간 7일)
지금 바로 교보eBook의 다양한 콘텐츠를 이용해 보세요!