- Evaluation and Analysis of Scattered Radiation Dose according to Factors in General X-ray Examination
- 발행기관
- 대한방사선과학회(구 대한방사선기술학회)
- 저자명
- 정동경(Dong-Kyung Jung) 박명환(Myeong-Hwan Park) 서정민(Jeong-Min Seo)
- 간행물 정보
- 『방사선기술과학』제47권 제1호, 13~19쪽, 전체 7쪽
- 주제분류
- 의약학 > 방사선과학
- 파일형태
- 발행일자
- 2024.02.29
국문 초록
영문 초록
Control of scattered radiation is one of very important factors in the use of medical radiation. In general X-ray exam, the causes, measurement methods, and the kind of detectors of scattered rays within the radiation area are diverse. In this study, the dose of scattered ray was measured by changing the thickness of the polycarbonate phantom and the tube voltage. As a result of measurement of scattered radiation, the results show that the scattered dose significantly(p<.05) increased with growing of thickness of phantom in the tube voltage 40, 50 and 60 kVp(F(p)<.05, R2>64%). As tube voltage increased at all phantom thicknesses, the scattered dose also significantly(p<.05) increased(F(p)<.05, R2>69%). In cases where a significant correlation was shown, the coefficient of determination of more than 60% was shown in regression analysis. The results of this study can be used as data on scattered radiation dose according to the tube voltage and the object thickness in general X-ray imaging exam.